GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 72
There are two pulse counters: Treatment Pulse Counter (shown on the “next” screen)
and Total Pulse Counter (shown on the “Laser Variable” Mode Screen). Both
counters keep track of delivered pulses but only the Treatment Pulse Counter can be
reset. The Total Pulse Counter is used to keep track of the total number of laser
pulses delivered by the laser system. The Treatment Pulse Counter is used to keep
track of the number of pulses used in a treatment session. Both pulse counts exclude
pulses used during calibration.
The Treatment Pulse Count is reset to zero by pressing the PULSE COUNT button
for approximately 3 seconds. The system acknowledges the selection by responding
with a short beep tone and by setting the lower count value displayed within the
PULSE COUNT button to zero. The Total Pulse Counter (shown on the “Laser
Variable” Mode Screen) can not be reset to zero.
The calibration procedure is initiated by pressing the CALIBRATE button. Note that
the system will initiate the calibration procedure automatically if READY state is
entered, a calibration is required, and the trigger switch is depressed. The handpiece
must be fully inserted into the cal port during the calibration routine.
After calibration, the system is in READY state and the trigger switch is enabled. The
operator can switch between READY state and STANDBY state as needed without
recalibrating provided that the fluence parameters have not been changed and no
more than thirty minutes has elapsed since the last calibration.
The language select button displays messages in the selected language. Press the
down or up arrow to select the desired language.
Pulse Count/
Reset
SYSTEM COUNTERS
Calibrate Button
Language Select
Содержание GentleYAG
Страница 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Страница 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Страница 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Страница 125: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 125 Labeling Symbols Section 7 ...
Страница 133: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 133 Label 13 ...
Страница 137: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 137 5 A B C 6 ...
Страница 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Страница 145: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 145 Accessory List Section 8 ...
Страница 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...