GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 108
Test #2 - Cryogen Coverage
:
Purpose
:
To verify the proper spray duration required to fill the distance gauge ring.
(Note: the distance gauge ring is larger than the spot size marking). Table 6-
1 shows the available distance gauges for the GentleYAG laser system.
GentleYAG Distance Gauge
Part Number
1.5mm
7122-00-3556 (small)\
7122-00-3561 (large)
3mm
7122-00-3535 (small)
7122-00-3534 (large)
8mm
7122-00-3502
10mm
7122-00-3503
12mm
7122-00-3109
15mm
7122-00-3504
18mm
7122-00-3505
Note: The part numbers shown for the 8 mm to 18 mm
distance gauges are for packages of 5 distance gauges.
Table 6-1
Available Distance
Gauges
Содержание GentleYAG
Страница 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Страница 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Страница 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Страница 125: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 125 Labeling Symbols Section 7 ...
Страница 133: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 133 Label 13 ...
Страница 137: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 137 5 A B C 6 ...
Страница 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Страница 145: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 145 Accessory List Section 8 ...
Страница 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...