pin is “ground” on your particular power plug, consult an Electrician for help. The
following procedure verifies ground continuity:
•
Using the Ohms setting of a Volt-Ohm meter, set the scale to "x1". Measure the
resistance between the plug's ground pin and any unpainted conductive surface
on the laser chassis. This reading must fall between 0 - 0.1 Ohms.
•
A battery & light or battery and buzzer combination may be alternatively used to
verify a ground connection between any unpainted conductive surface and the
plug's ground pin if an Ohmmeter is not available. An adequate ground
connection will be indicated by illumination of the light or sounding of the
buzzer.
Before installation of the
GentleYAG
Laser System, the intended site must be
prepared as described in this section. The site must have sufficient space to
accommodate the laser system, must provide the proper electrical power
configuration and receptacles, and must meet the additional environmental
specifications.
NOTE
Installation of the
GentleYAG
Laser System must be performed by a Candela
Service Representative. Following installation, a Candela Clinical Consultant
must instruct designated personnel on the basic operation and care of the laser.
An in-depth clinical training is required of a physician to become proficient in
the use of the
GentleYAG
Laser System.
NOTE
Treatment room areas associated with the use of cryogen require special
precautions. Refer to the Chemical Hazards paragraphs in section 2 of this
manual and the MSDS sheet (8501-00-1701) for further information.
ENVIRONMENTAL
REQUIREMENTS
Содержание GentleYAG
Страница 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Страница 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Страница 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Страница 125: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 125 Labeling Symbols Section 7 ...
Страница 133: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 133 Label 13 ...
Страница 137: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 137 5 A B C 6 ...
Страница 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Страница 145: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 145 Accessory List Section 8 ...
Страница 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...