Test #3 Cryogen Air Bubble Detection
Purpose:
To ensure that that air bubbles in the cryogen line are detected and the
associated “fault” message is displayed on the system.
Important: This procedure must be done with each GentleYAG
handpiece delivery system in your possession. The subsystem being
tested is inside of each delivery system, not in the laser system.
Procedure
1
Insure that the installed DCD canister is not empty.
Warning: Put on appropriate laser safety glasses
2.
Turn on the system and allow the “WARM UP” to
complete.
3.
With the Electrical Connector installed, disconnect the
Cryogen Connector (see Figure 3).
Содержание GentleYAG
Страница 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Страница 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Страница 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Страница 125: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 125 Labeling Symbols Section 7 ...
Страница 133: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 133 Label 13 ...
Страница 137: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 137 5 A B C 6 ...
Страница 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Страница 145: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 145 Accessory List Section 8 ...
Страница 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...