USB communications: use the EC100 USB cable to connect a PC to the base of the
EC100 enclosure.
o
RS-485 communications: use cable CABLE3TP-L for lengths less than 500 ft. The
connector on the EC100 panel labeled RS-485 displays which terminals are for
receiving and transmitting.
o
Analog output: use cable CABLE4CBL-L or CABLE2TP-L. Again, the customer
specifies the length of this cable at time of order. The connector labeled Analog
Outputs on the EC100 panel indicates where each wire should be connected (CO
2
,
H
2
O, and two ground connections).
7. Wire power and ground (power reference) cable CABLEPCBL-L to the EC100. Feed the
cable through one of the cable port openings in the base of the EC100 enclosure and
attach the ends into the green EC100 power connector. Plug the connector into the female
power connector on the EC100 panel. Ensure that the power and ground ends are going to
the appropriate terminals labeled 12V and ground, respectively.
8. Connect the power cable to a power source. The power and ground ends may be wired to
the 12V and G ports, respectively, of a Campbell Scientific data logger or to another 12 VDC
source.
9. Once power is applied to the EC100, three status LEDs on the EC100 panel will illuminate.
The power LED is green and the sonic and gas LEDs are orange until the unit has warmed
up and is ready to make measurements at which time the LEDs become green. If, after
several minutes, the LEDs turn red, a diagnostic flag has been detected. Check the
individual diagnostic bits to determine the specific fault. Diagnostics may be monitored
using the Status Window of
ECMon
, the user interface software included with the IRGASON
(see
(p. 19)), or with a data logger (see
Data collection and data processing
The diagnostics may reveal that the unit needs service such as cleaning the optical windows
on the gas analyzer, clearing the sonic transducers of ice or debris, etc. (see
(p. 31)).
7. Settings
Operation of the IRGASON can be customized by changing settings. Factory defaults work well
for most applications, but the user may adjust the settings with a PC using either the Campbell
Scientific
ECMon
software (see
(p. 24)) or
Device Configuration Utility
(see
(p. 26)), or with a data logger using the
EC100Configure()
instruction
(see
(p. 49)).
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
19
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...