After an extended period of time in the field, the IRGASON gas analyzer may need to undergo
factory calibration again to ensure valid measurements. When recalibration is deemed necessary,
contact Campbell Scientific.
The IRGASON sonic anemometer also undergoes an initial factory calibration. Usually this
calibration remains valid unless a transducer fails or damage to the instrument leads to a change
in geometry. The sonic anemometer requires calibration when:
l
it develops a wind offset greater than the specification
l
it sets diagnostic flags under dry conditions with little to no wind and with no obstruction in
the ultrasonic paths
The wind offset is tested by creating a zero-wind environment. This is best done in a laboratory
setting with HVAC vents closed or covered to reduce air currents and by encircling the mounted
sensor with a large plastic bag (for example, an unused refuse bag), using caution not to block
the sonic paths. Once the IRGASON is connected to an EC100 and powered on, wind offsets may
be viewed by connecting the EC100 to a PC and using
ECMon
to graph u
x
, u
y
, and u
z
wind
components (see
(p. 24)). In this zero-wind environment, u
x
and u
y
should be less than ±
8 cm∙s
-1
(± 0.08 m∙s
-1
) and u
z
should be less than ± 4 cm∙s
-1
(± 0.04 m∙s
-1
). If recalibration is
deemed necessary, contact Campbell Scientific.
10. Data collection and data
processing
Data from the IRGASON is collected through the EC100 and then archived onto a data logger. A
common instrument configuration is to program a data logger to retrieve and collect raw data
from the IRGASON, to be used for post processing, for which various programs have been
developed.
More recently, programs have been developed that efficiently record and correctly process data
from instruments such as the IRGASON, as well as compile them with data from other,
complementary instruments. Campbell Scientific has developed a program, the
EasyFlux DL
, that
both records and processes raw data from the IRGASON to provide useful measurements
immediately. An overview of both approaches is given in the sections below.
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
42
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...