Ambient relative humidity (in humidity greater than 80% heaters will try to maintain
internal temperature 2 degrees warmer than ambient)
l
CO
2
signal level (1 min average CO
2
signal level; below 0.7 will cause the heater to turn on
maximum power until the signals recover)
l
Average slope of the CO
2
signal level over 1 min
l
Standard deviation of the CO
2
signal over 1 min
When enabled, this setting applies a voltage between 0 and 4000 mV to heaters near the optical
windows of the analyzer. Heated windows inhibit the formation of condensation, such as dew
and frost, and help the analyzer recover more quickly when precipitation has blocked the optical
path. Use this setting with caution. Excessive heating may lead to density fluctuations and
subsequent flux measurement errors. More information about heater control, and specifically
using a data logger for control, may be found at the end of
7.14
ECMon
Settings for the IRGASON are easily verified or changed using the Windows PC support software
ECMon
(Eddy Covariance Monitor), which is found at
.
Launch
ECMon
, and connect the EC100 electronics to the PC with the included EC100 USB cable.
The USB connection for the EC100 electronics is found on the base of the enclosure (see
(p. 17)). Once connected, select the appropriate communications port in the
ECMon
main
page and click Connect (see
(p. 25)). Next, click on the Setup button. All of the
preceding settings are now available for editing (see
(p. 26)).
Besides changing settings,
ECMon
is a useful tool for common tasks such as:
l
Monitoring real-time data from the IRGASON from the main window
l
Performing manual zero and span of the instrument (
l
Troubleshooting and monitoring diagnostics using the Status window.
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
24
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...