The wicks are constructed with a UV-resistant fabric and should function over long time periods.
Nevertheless, they should be inspected periodically. Check for contamination from dust, pollen,
pitch or other debris. If needed, wash them in warm water with mild detergent or replace them.
When performing the zero-and-span procedure, the wicks should be removed before installing
the zero-and-span shroud.
Remove the wicks during the winter. The wicks will accumulate snow or freezing rain, resulting in
blockage of the optical path. Besides the wicks, controlling the heaters in the snouts is another
method the user may consider in preventing data loss during precipitation and condensation
events. The heaters can be controlled statically using the Setup window in
ECMon
or dynamically
using a data logger (see
(p. 49) for more details).
FIGURE 9-3. Proper location of the gas analyzer top wick (left) and bottom wick (right)
9.3 Gas analyzer window cleaning
The windows of the analyzer should be cleaned if the signal strength for CO
2
or H
2
O drops below
0.7 or 70% of the original value. These values may be monitored in the output data, or they can
be viewed with
ECMon
. To clean the windows, use isopropyl alcohol (2-propanol/isopropanol)
and a cotton swab or a non-scratching tissue or cloth. Verify that following cleaning the signal
strengths are restored to values close to 1.0.
9.4 Zero and span
As is the case with all optical instruments, the IRGASON gas analyzer measurements may drift
slightly with exposure to natural elements. Therefore, routine maintenance requires a periodic
zero-and-span procedure. The frequency with which this should be done is heavily dependent
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
34
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...