Table 10-1: Output modes for EC100 instruction
Output mode
Data field
Description
Units
0, 1, 2
1
Ux
m/s
2
Uy
m/s
3
Uz
m/s
4
Sonic Temperature
ºC
5
Sonic Diagnostic Flag
6
CO
2
Density
mg/m
3
7
H
2
O Density
g/m
3
8
Gas Diagnostic Flag
1, 2
9
Air Temperature
ºC
10
Air Pressure
kPa
11
CO
2
Signal Strength
nominally 0.0 ≤
strength ≤1.0
12
H
2
O Signal Strength
nominally 0.0 ≤
strength ≤1.0
2
13
CO
2
Density from fast-
response temperature
mg·m
–3
*This additional CO
2
density output is available in EC100 OS version 7.01 or later. It uses humidity-corrected sonic
temperature instead of ambient temperature measured by the EC100 temperature probe in the conversion of
absorption measurements to CO
2
density. Using the sonic anemometer’s fast-response temperature
measurements, compensates for spectroscopic effects during high sensible heat flux regimes as explained in
Helbig
et al.
(2016).
NOTE:
An open-path gas analyzer connected to an EC100 running an OS version less than 7.01, the
13
th
data field/element is unused, If a closed-path gas analyzer (for example, an EC155) is
connected to the EC100, this data field/element will be the sample cell pressure differential.
As shown in
(p. 47), all output modes give two diagnostic values: Sonic Diagnostic Flag
and Gas Diagnostic Flag. The values contain a bit field with each bit representing a monitored
condition. When a certain condition is detected, the corresponding bit is set. The value remains
set until the event that caused the condition is no longer present.
(p. 48) and
(p. 48) describe the bits in the Sonic Diagnostic Flag and the Gas Diagnostic Flag, respectively.
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
47
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...