7.9 Temperature sensor
This setting configures the EC100 electronics to integrate with the IRGASON. The IRGASON
measures ambient air temperature using a thermistor temperature sensor. With Auto-Select
selected, the EC100 will detect that the IRGASON is connected and will report ambient air
temperature measurements from the thermistor.
To diagnose problems with the temperature measurement, a fixed temperature value may be
used, or the temperature sensor may be selected manually.
7.10 Fixed temperature value
If the Temperature Sensor setting is None, the IRGASON will use the value of this setting for the
sample temperature. This mode is intended for troubleshooting only. In normal operation,
Temperature Sensor is set to Auto-Select, and this setting is not used.
7.11 Pressure sensor
This setting identifies which barometric pressure sensor is used. The EC100 always includes the
basic barometer, but it may be ordered with the optional enhanced barometer (model CS106).
This setting defaults to EC100 Enhanced if the CS106 is ordered and to EC100 Basic otherwise.
There are two other possible settings for Pressure Sensor. First, the user may supply a different
pressure sensor. In this case, the setting should be changed to User Supplied, with the
appropriate values for gain and offset entered (see below).
The final option is to select None for Pressure Sensor. The EC100 will use a fixed (see below) value
for pressure. This mode is intended for troubleshooting only.
7.11.1 Pressure gain
If Pressure Sensor is set to User Supplied, this setting gives the gain factor (kPa/V) used to
convert measured voltage to pressure. Normally, Pressure Sensor is set to EC100 Basic or EC100
Enhanced, and this setting is not used.
7.11.2 Pressure offset
If Pressure Sensor is set to User Supplied, this setting gives the offset (kPa) used to convert
measured voltage to pressure. Normally, Pressure Sensor is set to EC100 Basic or EC100
Enhanced, and this setting is not used.
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
22
Содержание IRGASON
Страница 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Страница 72: ...D 1 Molecular sieve Type 13X IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 68 ...
Страница 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Страница 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Страница 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Страница 76: ...D 2 Magnesium perchlorate IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 72 ...
Страница 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Страница 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Страница 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Страница 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Страница 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Страница 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Страница 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Страница 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Страница 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Страница 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...