SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
7-14
FIGURE 7.13-1. A Vibrating Wire Sensor
The following calculations are based on using a
Geokon model 4500 Vibrating Wire sensor. An
individual multiplier and offset must be
calculated for each sensor used in a system.
MULTIPLIER
The fundamental equation relating frequency to
pressure is
P = -F
x
G + B where
P = pressure, PSI
G = the Gage Factor obtained from the
sensors calibration sheet in PSI/digit.
The units of a digit are Hz
2
(10
-3
).
B = offset
F
x
= f
2
Hz
2
(10
-3
), where f is frequency.
Instruction 28 measures period, T, of the
vibrating wire in milliseconds (ms) and returns a
measured value, X, of
X = 1/(T
2
(ms)
2
) = f
2
(10
-6
)Hz
2
A multiplier of -1000 in Instruction 28 converts
the measurement to digits, as shown below.
-F
x
= -X(-10
3
) = -f
2
(10
-3
)Hz
2
To calculate the multiplier, convert Geokon's
gage factor, G, to the desired units (i.e., feet of
water per digit) and multiply by -1000 digits/kHz
2
.
TEMPERATURE CORRECTION
The temperature correction is applied as
follows.
P
T
= P + C
∗
(t
1
- t
0
), where
P
T
= Pressure corrected for temperature,
°
C
C
= Temperature coefficient, PSI/
°
C
(from Geokon calibration sheet)
t
0
& t
1
= Initial and current temperatures,
°
C.
The temperature coefficient, C, must be
converted to units compatible with the gage
factor, G.
WELL MONITORING EXAMPLE
In this example the vibrating wire sensor is used
to monitor water table height (Figure 7.13-2).
The desired data is the distance from the lip of
the well to the water surface. The sensor is
vented to atmosphere to eliminate
measurement errors due to changes in
barometric pressure. The water level is
expected to stay within 40 to 80 feet of the lip so
the 50 psi pressure sensor is placed
approximately 100 feet below the lip of the well.
The calibration data from Geokon is provided in
Table 7.13-1.
TABLE 7.13-1 Calibration Data for
Sensor 3998
Gage Factor
Temp. Coeff.
(psi/digit)
(psi/
°
C)
0.0151
-0.0698
Содержание CR510
Страница 1: ...CR510 DATALOGGER OPERATOR S MANUAL REVISION 2 03 COPYRIGHT c 1986 2003 CAMPBELL SCIENTIFIC INC ...
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 10: ...This is a blank page ...
Страница 44: ...SECTION 1 FUNCTIONAL MODES 1 14 This is a blank page ...
Страница 56: ...SECTION 3 INSTRUCTION SET BASICS 3 8 ...
Страница 104: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 8 This is a blank page ...
Страница 128: ...SECTION 10 PROCESSING INSTRUCTIONS 10 8 This is a blank page ...
Страница 144: ...SECTION 12 PROGRAM CONTROL INSTRUCTIONS 12 10 This is a blank page ...
Страница 172: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 10 This is a blank page ...
Страница 176: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 184: ...This is a blank page ...
Страница 188: ...APPENDIX D DATALOGGER INITIATED COMMUNICATIONS D 4 This is a blank page ...
Страница 196: ...APPENDIX F MODBUS ON THE CR10 X AND CR510 F 4 This is a blank page ...
Страница 197: ...APPENDIX G TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Страница 198: ...This is a blank page ...
Страница 200: ...This is a blank page ...
Страница 206: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Страница 222: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Страница 238: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Страница 240: ...This is a blank page ...