![Campbell CR510 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/campbell/cr510/cr510_operators-manual_489094078.webp)
SECTION 6. 9-PIN SERIAL INPUT/OUTPUT
6-8
IF NOTHING HAPPENS
If the CR510 is connected to the SC32A RS232
interface and a modem/terminal, and an "
∗
" is
not received after sending carriage returns:
1.
Verify that the CR510 has power AT THE
12V AND GROUND INPUTS, and that the
cables connecting the devices are securely
connected.
2.
Verify that the port of the modem/terminal is
an asynchronous serial communications
port configured as DTE (see Table 6.7-2).
The most common problems occur when
the user tries to use a parallel port, or
doesn't know the port assignments, i.e.
COM1 or COM2. IBM, and most
compatibles come with a Diagnostic disk
which can be used to identify ports, and
their assignments. If the serial port is
standard equipment, then the operators
manual should give you this information.
3.
Verify that there is 5 volts between the
CR510 5V and G terminals. Call Campbell
Scientific technical support if the voltage is
less than 4.8 volts.
Some serial ports, e.g., the Super Serial Card
for Apple computers, can be configured as DTE
or DCE with a jumper block. Pin functions must
match Table 6.7-2.
If you are using a computer to communicate
with the datalogger, communication software
must be used to enable the serial port and to
make the computer function as a terminal. The
port should be enabled for 300, 1200, or 9600
baud, 8 data bits, 1 stop bit, and no parity.
Campbell Scientific's GraphTerm, PC208E,
PC208W, and TCOM provide this function.
If you are not sure that your computer/terminal
is sending or receiving characters, there is a
simple way to verify it. Make sure that the
duplex is set to full. Next, take a paper clip and
connect one end to pin 2, and the other end to
pin 3 of the serial port. Each character typed on
the keyboard will be displayed only if
transmitted from the terminal on pin 2, and
received on pin 3 (if duplex is set to half, the
character will be displayed once if it is not
transmitted, or twice if it is transmitted).
IF GARBAGE APPEARS
If garbage characters appear on the display,
check that the baud rate is supported by the
CR510. If the baud rate is correct, verify that
the computer/terminal is set for 8 data bits, and
no parity. Garbage will appear if 7 data bits and
no parity are used. If the computer/terminal is
set to 8 data bits and even or odd parity,
communication cannot be established.
Содержание CR510
Страница 1: ...CR510 DATALOGGER OPERATOR S MANUAL REVISION 2 03 COPYRIGHT c 1986 2003 CAMPBELL SCIENTIFIC INC ...
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 10: ...This is a blank page ...
Страница 44: ...SECTION 1 FUNCTIONAL MODES 1 14 This is a blank page ...
Страница 56: ...SECTION 3 INSTRUCTION SET BASICS 3 8 ...
Страница 104: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 8 This is a blank page ...
Страница 128: ...SECTION 10 PROCESSING INSTRUCTIONS 10 8 This is a blank page ...
Страница 144: ...SECTION 12 PROGRAM CONTROL INSTRUCTIONS 12 10 This is a blank page ...
Страница 172: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 10 This is a blank page ...
Страница 176: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 184: ...This is a blank page ...
Страница 188: ...APPENDIX D DATALOGGER INITIATED COMMUNICATIONS D 4 This is a blank page ...
Страница 196: ...APPENDIX F MODBUS ON THE CR10 X AND CR510 F 4 This is a blank page ...
Страница 197: ...APPENDIX G TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Страница 198: ...This is a blank page ...
Страница 200: ...This is a blank page ...
Страница 206: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Страница 222: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Страница 238: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Страница 240: ...This is a blank page ...