TD ADDENDUM—OVERVIEW
AD-OV-7
The program to make the measurements and send the desired data to Final Storage has been
entered. The program is complete. The clock must now be set so that the date and time tags
are correct. (Here the example reverts back to the key by key format.)
Key
Display
Explanation
*5
00:21:32
Enter *5 Mode. Clock running but not set correctly.
A
05:01.01
Advance to month-day (MMDD).
1004
05:1004
Key in MMDD (Oct 4 in this example).
A
05:1990
Enter and advance to location for year.
1994
05:1994
Key in year.
A
05:00:21
Enter and advance to location for hours and minutes
(24 hr. time).
1324
05:1324
Key in hrs.:min. (1:24 PM in this example).
A
05:27.250
Key in seconds
30
05:30
A
13:24:30
Clock set and running. Changes made when A was
pressed.
*0
LOG 1
Exit *5, compile Table 1, commence logging data.
OV5.3 EDITING AN EXISTING PROGRAM
When editing an existing program in the
CR10X, entering a new instruction inserts the
instruction; entering a new parameter replaces
the previous value.
To insert an instruction, enter the program table
and advance to the position where the
instruction is to be inserted (i.e., P in the data
portion of the display) key in the instruction
number, and then key A. The new instruction
will be inserted at that point in the table,
advance through and enter the parameters.
The instruction that was at that point and all
instructions following it will be pushed down to
follow the inserted instruction.
An instruction is deleted by advancing to the
instruction number (P in display) and keying #D
(Table 4.2-1).
To change the value entered for a parameter,
advance to the parameter and key in the correct
value then press A. Note that the new value is
not entered until A is keyed.
OV6. DATA RETRIEVAL OPTIONS
Data is retrieved over some form of
telecommunications link, whether it be RF
(radio), telephone, short haul modem, coaxial
cable (mulitdrop) , or direct link. The table data
operating system does not support on-line
output to peripheral storage devices (see Figure
OV 6.1-1).
The retrieval of data does NOT erase those
data from the Data Storage Tables in Final
Storage. The data remains in the table ring
memory until:
The are written over by new records of data.
(Section 2.1)
Input Storage memory is reallocated
(Section 1.5)
The datalogger program is changed and
compiled.
Power to the datalogger is turned off.
Содержание CR510
Страница 1: ...CR510 DATALOGGER OPERATOR S MANUAL REVISION 2 03 COPYRIGHT c 1986 2003 CAMPBELL SCIENTIFIC INC ...
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 10: ...This is a blank page ...
Страница 44: ...SECTION 1 FUNCTIONAL MODES 1 14 This is a blank page ...
Страница 56: ...SECTION 3 INSTRUCTION SET BASICS 3 8 ...
Страница 104: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 8 This is a blank page ...
Страница 128: ...SECTION 10 PROCESSING INSTRUCTIONS 10 8 This is a blank page ...
Страница 144: ...SECTION 12 PROGRAM CONTROL INSTRUCTIONS 12 10 This is a blank page ...
Страница 172: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 10 This is a blank page ...
Страница 176: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 184: ...This is a blank page ...
Страница 188: ...APPENDIX D DATALOGGER INITIATED COMMUNICATIONS D 4 This is a blank page ...
Страница 196: ...APPENDIX F MODBUS ON THE CR10 X AND CR510 F 4 This is a blank page ...
Страница 197: ...APPENDIX G TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Страница 198: ...This is a blank page ...
Страница 200: ...This is a blank page ...
Страница 206: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Страница 222: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Страница 238: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Страница 240: ...This is a blank page ...