MC-II Flow Analyzer
21
3.
After the diagnostics routine is complete, the upper display will show
the prompt
tot.Eng.
Press
INCREMENT
until
USEr
appears on the
lower display. Press
ENTER/STEP
to confirm the selection.
4.
The upper display will show
Ent.diV
, prompting you for the divisor.
The lower display will show the previously entered meter factor (fac-
tory default is 230.00). The right-most digit (the ones position) will
blink, indicating it is the digit currently selected for editing.
a.
Remove the decimal point by pressing
DEC. POINT
repeatedly.
b.
Press
INCREMENT
until 7 is displayed in the ones place.
(If you accidentally pass the desired digit, continue to press
IN-
CREMENT
until that digit is displayed again.) Press
ENTER/STEP
to proceed to the next digit (tens position).
c.
Press
INCREMENT
until 3 is displayed. Press
ENTER/STEP
to
proceed to the next digit (hundreds position).
d.
Press
INCREMENT
until 4 is displayed. Press
ENTER/STEP
to
proceed to the next digit (thousands position).
e.
Press
INCREMENT
until 6 is displayed. Press
ENTER/STEP
to
proceed to the next digit (ten thousands position).
f.
Press
INCREMENT
until 8 is displayed. Press
ENTER/STEP
to
proceed to the next digit (one hundred thousands position).
g.
Press
INCREMENT
until 7 is displayed. Press
ENTER/STEP
to
confirm the entry of the divisor.
Enter the calculated divi-
sor.
Factory
default or
last divisor
entered
a.
b.
c.
d.
e.
f.
g.
Underline denotes the flash-
ing digit to be changed.
Содержание NuFlo MC-II
Страница 1: ...NUFLO MC II FlowAnalyzer User Manual Manual No 100079666 Rev OE ...
Страница 6: ...MC II Flow Analyzer 2 Figure 1 Nomenclature exterior ...
Страница 7: ...MC II Flow Analyzer 3 Figure 2 Nomenclature interior ...
Страница 32: ...MC II Flow Analyzer 28 ...
Страница 35: ...Figure A2 Field wiring for dry contact pulse output option non hazardous areas MC II Flow Analyzer 31 ...
Страница 36: ...32 Figure A3 Field wiring for open collector pulse output option non hazardous areas MC II Flow Analyzer ...
Страница 37: ...Figure A4 Field wiring for opto isolated pulse output option non hazardous areas MC II Flow Analyzer 33 ...
Страница 38: ...MC II Flow Analyzer 34 ...
Страница 44: ...40 MC II Flow Analyzer ...
Страница 45: ......
Страница 46: ......