MC-II Flow Analyzer
12
Bypass Security Password
Forget your password? With the MC-II, you can have your access restored
quickly and easily with a one-time bypass code.
Press RESET
simultaneously to display
the bYPASS menu.
TOTAL
RATE
Call 1-800-654-3760 to
request a one-time
security bypass code.
To obtain this bypass code, perform the following steps:
1.
Press and hold
ACCESS
and
RESET
simultaneously for 3 seconds,
then release. The word “bYPASS” will appear in the upper display,
and a number will appear in the lower display.
2.
Record this number (it will remain on the display panel only for a few
seconds).
3.
Call a Cameron (Measurement Systems Division) representative, and
give him/her the number and the serial number of the MC-II. After
verifying your identity, he/she will provide you with a one-time by-
pass code.
4.
Enter the bypass code in place of your password at the
CodE
prompt.
Important—Prior to exiting the Calibration menu, the
CodE
menu is
displayed again, and zeroes will appear in the bottom display. Select a
password using the
INCREMENT
key, and press
ENTER/STEP
to save
your password.
To re-enter a numerical calibration setting before you have selected
and saved the last digit, press and hold
RESET
for 3 seconds, then
release. The right-most digit will resume blinking, allowing you to
enter the entire number again.
To retain a numerical calibration setting and advance to the next
menu without pressing
ENTER/STEP
repeatedly, press and hold
ACCESS
for 3 seconds, then release. This shortcut applies only to
numerical fields. You must press
ENTER/STEP
to select all other
entries.
If the MC-II is left in the Calibrate mode for longer than 1 minute, it
will time-out and return to the Run mode. Data that has been saved
with the
ENTER/STEP
key will be saved. However, data that was
selected but not entered will be lost.
Data Entry Tips
000000
Содержание NuFlo MC-II
Страница 1: ...NUFLO MC II FlowAnalyzer User Manual Manual No 100079666 Rev OE ...
Страница 6: ...MC II Flow Analyzer 2 Figure 1 Nomenclature exterior ...
Страница 7: ...MC II Flow Analyzer 3 Figure 2 Nomenclature interior ...
Страница 32: ...MC II Flow Analyzer 28 ...
Страница 35: ...Figure A2 Field wiring for dry contact pulse output option non hazardous areas MC II Flow Analyzer 31 ...
Страница 36: ...32 Figure A3 Field wiring for open collector pulse output option non hazardous areas MC II Flow Analyzer ...
Страница 37: ...Figure A4 Field wiring for opto isolated pulse output option non hazardous areas MC II Flow Analyzer 33 ...
Страница 38: ...MC II Flow Analyzer 34 ...
Страница 44: ...40 MC II Flow Analyzer ...
Страница 45: ......
Страница 46: ......