MC-II Flow Analyzer
16
Enable or disable the
Preset Total option.
7.
The prompt
SEt.tot
will appear in the top display, and
no
(default
setting) will appear in the lower display.
a.
If you do not need to preset a total, press
ENTER/STEP
to by-
pass this menu selection.
The prompt
SAVing
will appear in the lower display while the
calibration settings are saved to nonvolatile memory. The MC-II
will automatically return to the Run mode.
b.
To preset a total, press
INCREMENT
to toggle the bottom
display from
no
to
YES
. Press
ENTER/STEP
to accept the
selection. A series of zeros will appear in the bottom display, and
the right-most digit will begin flashing, allowing you to enter a
total. Enter the digits for the preset value, right to left, by press-
ing
INCREMENT
repeatedly to select the appropriate digit,
then pressing
ENTER/STEP
to accept the selection. Repeat this
process for all digits. When the last digit has been entered and
ENTER/STEP
is pressed, the calibration is complete.
The prompt
SAVing
will appear in the lower display while the
calibration settings are saved to nonvolatile memory. The MC-II
will automatically return to the Run mode and begin totalizing
volume beginning with the new preset total.
Important—If the unit of measure selected for the flow rate results in a
rate that contains too many digits to be displayed in the LCD window,
the MC-II will generate an overrun error. A prompt,
oU run
, will appear
in the bottom display. To correct this error, press and hold
ACCESS
for 3
seconds, then release, to reenter the Calibrate mode; then select a larger
unit of measure for the flow rate.
Example calibration procedures are provided on the following pages for
these basic applications:
•
liquid measurement with preprogrammed units of measure (page 17)
•
liquid measurement with user-defined units of measure (page 20)
•
gas measurement (page 24)
Содержание NuFlo MC-II
Страница 1: ...NUFLO MC II FlowAnalyzer User Manual Manual No 100079666 Rev OE ...
Страница 6: ...MC II Flow Analyzer 2 Figure 1 Nomenclature exterior ...
Страница 7: ...MC II Flow Analyzer 3 Figure 2 Nomenclature interior ...
Страница 32: ...MC II Flow Analyzer 28 ...
Страница 35: ...Figure A2 Field wiring for dry contact pulse output option non hazardous areas MC II Flow Analyzer 31 ...
Страница 36: ...32 Figure A3 Field wiring for open collector pulse output option non hazardous areas MC II Flow Analyzer ...
Страница 37: ...Figure A4 Field wiring for opto isolated pulse output option non hazardous areas MC II Flow Analyzer 33 ...
Страница 38: ...MC II Flow Analyzer 34 ...
Страница 44: ...40 MC II Flow Analyzer ...
Страница 45: ......
Страница 46: ......