IN-SERVICE TRAINING OUTCOMES
Once your Microdot® Xtra Meter in-service is complete, Health Care Professionals
should be able to:
1. Locate and explain the following
components of the Microdot Xtra Meter:
Eject Button
Test Strip Port
Battery Compartment
Serial Number
Customer Service Number
2. Locate and explain the label
information on the Microdot Xtra Test
Strip Vial:
Discard Date
Lot Number
Expiration Date
Control Ranges
Storage Temperature Range
3. Set Time and Date of meter
4. Identify three parts of the
Microdot Xtra Test Strips
Top Edge
Test Pad
Contact Bars
5. Identify the following parts of the
Microdot Control Solution Bottles:
Low vs. High
Discard Date
Lot Number
Expiration Date
Storage Temperature Range
6. Properly insert the Microdot Xtra Test Strip
7. Perform control solution tests
8. Document and maintain Microdot Xtra
Quality Control Records
9. Obtain a blood sample
10. Perform a blood test
11. Identify meter result range
12. Access meter memory
13. Change the battery
14. Identify and resolve error readings
Chapter 6
14
Cambridge Sensors USA, LLC
METER SPECIFICATIONS
Sample Size
Total Test Time
Sample Application
Hematocrit Range
Measurement Range
Interference
Strip Removal
Coding
Accuracy of Strip/Meter System
Precision of Strip
Shelf Life of Meter
Operating Humidity range
Operating Temperature
Type of Glucose Result
Strip Vial Packing Size
Typical Control Solution Ranges
when used with Strips/Meter
Strip Shelf Life after Opening
Strip Vial
Altitude
Module Size
Weight
Battery Life
Glucose Units
Memory
Power Source
Automatic Shut-off
Warranty
0.6 Microliters (600 Nanoliters)
10 Seconds
Strip is placed in the meter and sample
wicks into strip
30 - 50%
20 to 525 mg/dL
No interference from over 20 interfering substances
Strip ejection by push button
Auto-code
At least +/- 20% relative to YSI in clinical trials
Correction Coefficient of Regression > 0.969
The strip variation is not greater than 6.4%
Approx. 5 years
10 - 90%; storage 90% max (non-condensing)
50˚ - 104˚F (10˚ - 40˚C)
Plasma equivalent
50 strips per vial
Printed on vial label
3 months
Plastic with desiccant sleeve vial
To 10,000 ft (Target)
65 mm x 85 mm x 16 mm
Approx. 60 grams / 2 ounces
1,500 tests or about 1 year at three tests per day
mg/dL or mmol/L
Total of 500 glucose results and insulin-input data
One replaceable 3V lithium battery
(CR2032 or equivalent)
One minute after last user action
1 Year Manufacturer Warranty
Chapter 5
13
Cambridge Sensors USA, LLC
®