background image

“Haiim” Vacuum-assisted blood 

  collection system 

Please read this instruction manual carefully

before operating this product.

User Manual

A.【Product Description】

This  product,  “Haiim”  Vacuum-assisted  blood  collection  system  (WH-001),  is 
composed of two parts: 1) a main device (HD-001) and 2) a single-use disposable 
cassette  (HC-001).  The  main  device  includes  a  pressure  control  system  that 
generates negative pressure to create a vacuum effect for collecting blood. The 
intended purpose of the product is limited to IVD use only. When operating this 
product, it is necessary to use a blood lancet with a needle (diameter ≥ 0.8 mm, 
gauge number ≤ 21G) or a blade (width ≥ 1.5 mm) and a micro-collection tube 
(the  information  of  compatible  micro-collection  tubes  is  provided  in  Section  E, 
Cassette (HC-001) General Specifications).

(1) Main Device

(2) Cassette

(3) Micro-

collection tube

(4) Lancet

White light:

Green light:

Blue light:

2. Cassette (HC-001)

2.1 Blood Collection Entry: an open end of the cassette on where a fingertip 
should be placed.
2.2  Main  Device  Connector:  a  connector  for  connecting  to  the  Cassette 
Connector.
2.3 Tube Connector: a connector with a specific design angle for connecting to 
a commercial micro-collection tube recommended in this user manual.
2.4 O-ring.
2.5  Blood  Collection  Hole:  the  top  entry  of  the  capillary  structure  of  the 
cassette for receiving blood specimens from a fingertip.

2.3 Tube Connector

 2.1 Blood Collection Entry

2.2 Main Device

Connector

H.【Preceding Operation】

Before starting, please prepare the following items:

B.【Intended Use】

C.【Contraindications】

“Haiim” Vacuum-assisted blood collection system is intended to use a vacuum to 
collect  capillary  blood  from  a  puncture  site.  The  product  is  composed  of  a 
cassette  able  to  connect  with  a  single-use  micro-collection  tube  and  a  main 
device providing a pressure control function. When assembled and activated, the 
product collects blood from the puncture site to the micro-collection tube. This 
product is for professional use only.

1. Diseases that are associated with blood coagulation disorders or need to take 
any anti-coagulation medication/treatment may cause abnormal blood loss.
2.  Areas  of  skin  infection  or  skin  conditions  like  cellulitis  or  abscess  should  be 
avoided because of the risks of inoculating the blood with bacteria, viruses, and 
any substance that caused infection.

D.【Warnings and Precautions】

1 . P r o d u c t   S a f e t y

1.1 Before starting to use this product, please confirm if the standard voltage of 
the country or region where you are going to use the product is in the range of 
the technical specifications of the power adapter provided in Section E of this user 
manual.
1.2 Before starting to use this product, please verify that the product has complete 
packaging shown in Section F. If there is any missing part, please kindly return the 
whole product to where you purchased it from for a replacement.
1.3 Do not apply any physical impact to the main device and avoid it from falling.
1.4 This product can be used in a professional healthcare environment.
1.5 This product is not intended for use in residential environments and may not 
provide adequate protection to radio reception in such environments.
1.6 The product should not be used adjacent to or stacked with other equipment.
1.7 Do not insert anything into the connectorsʼ holes or the outer case gaps of the 
main device.
1.8 Do not place the product in a location accessible to children or unauthorized 
users.

3 . I n s t r u c t i o n s   f o r   S a f e   B l o o d   D r a w i n g

3.1 Before starting to use this product, please read this user manual carefully and 
other documents provided along with the product.
3.2 This product has been designed for fingertip blood collection. Please do not 
give it any other uses. 
3.3 Only medical or healthcare providers are eligible to operate this product in 
accordance with local regulations.
3.4 The main device can only be used with the cassettes. Please do not use any 
other goods similar to the cassettes.
3.5 The cassette is a single-use consumable. Please do not reuse it.
3.6 Do not use any cassette whose individual packaging is damaged or perforat-
ed to avoid any risks of contamination. 
3.7  Do  not  start  to  operate  this  product  until  the  cassette  has  been  correctly 
installed.
3.8  For  emergencies  or  accidents  during  blood  drawing,  please  refer  to  the 
system operation setting stated in Section E of this user manual.
3.9  Users  should  prepare  other  items  that  may  be  required  for  blood  drawing 
according to Section H (e.g., gloves, alcohol pads, etc.).
3.10 Since blood is a high-risk biological specimen, it is important to take proper 
precautions  every  time  when  performing  the  blood  collection  process.  Users 
shall avoid directly touching the wounds or any areas with blood.
3.11  Do  not  inject  the  collected  blood  specimen  back  into  anyone's  body  to 
avoid infection or further risks.
3.12 Usage of this product performing by a person unable to properly operate 
the product due to color blindness, blindness, or blurred vision shall be avoided.
3.13 Please comply with local authority regulations and use appropriate biologi-
cal waste disposal protocols to dispose of used cassettes and other consumables 
that have been used or that have been in contact with blood specimens. Regard-
ing the disposal of the main device, please follow your national requirements.

4 . C l e a n i n g   a n d   M a i n t e n a n c e   C o n s i d e r a t i o n s

4.1 Please avoid dust, dirt, or other contaminants enters the Cassette Connector 
of  the  main  device  (see  section  G,  1.4)  or  accumulated  on  the  main  device 
surfaces.
4.2 The main device shall be cleaned regularly, especially when it has been in 
contact with blood specimens, to prevent the risk of cross-contamination. For 
more details, please refer to the Section J of this user manual.
4.3 It is forbidden to disassemble or alter completely or partially both the main 
device and the cassette. If so, the functionality and safety of this product are not 
assured. 
4.4 If the product requires maintenance, please contact the customer service 
provided by Winnoz Technology, Inc.

E.【Items and Specifications】

“Haiim” Vacuum-assisted blood collection system

Main Device(HD-001)

 

General Specifications

F.【Product Images】

The complete packaging of “Haiim” Vacuum-assisted blood collection system 

is shown below: 

1. Main device (HD-001), including Power Adapter (GEM12I12-P1J)

EMC Standards

· Conducted Emission (IEC 60601-1-2:2014)

· Radiated Emission (IEC 60601-1-2:2014)

· Harmonic distortion (IEC 60601-1-2:2014)

· Voltage fluctuations and flicker (IEC 60601-1-2:2014)

· ESD (IEC 60601-1-2:2014)

· RS Radiated RF EM fields (IEC 60601-1-2:2014) 

· RS Proximity fields from RF wireless communications

  equipment (IEC 60601-1-2:2014)

· EFT (IEC 60601-1-2:2014)

· Surge (IEC 60601-1-2:2014)

· CS (IEC 60601-1-2:2014)

· PFMF (IEC 60601-1-2:2014)

· Voltage dips and interruptions (IEC 60601-1-2:2014)

Dimension

Weight

Recommended

Micro-collection

Tube

Specification

Compatible

Device
Packaging
Operating

Condition

Shelf Life

Width: 18 mm/Length: 32.5 mm/Height: 42 mm

(excluding the height of the assembled micro-

 collection tube)

4 g (Net weight)

1. BD Microtainer® Tube with BD Microgard™ Closure

    0.5 mL (Without gel)

2. EV Single-use Containers for Human Capillary Blood

    Specimen Collection 0.5 mL (Without gel)

It is not recommended to use a micro-collection tube

with gel.

“Haiim” Vacuum-assisted blood collection system. 

Main Device (HD-001)
Single package (non-sterile)
Temperature range:0~40℃

Humidity range: 30~75 % RH

3 years

Dimension

Weight

System

Operation

Setting

Power Adapter

Specifications

Compatible

Cassette

Operating

Conditions

Storage

Conditions

Width: 62 mm/Length: 123 mm/Height: 62 mm

219 g (net weight)
- Press the Start-Button (Section G, 1.3) to start.

- To stop the main device when functioning in case

  of emergency or accidents, press the Start-Button

  for one second.

- Automatic cycle time: 2 min
Model: GEM12I12-P1J

Input Rating: 100-240 VAC, 50-60 Hz, 0.4-0.2A 

Output Rating: 12 Vdc/1A 

Cable Length: 1.8 m

Weight: 118 g
“Haiim” Vacuum-assisted blood collection system

 Cassette (HC-001)
Temperature range:0~40℃

Humidity range: 30~75% RH
Temperature  range:0~40℃

Humidity range:30~75% RH

Avoid direct exposure to sunlight

Item Name

Model

Remarks

“Haiim” Vacuum-assisted 

blood collection system
User Manual

Main Device

Cassette
Power adapter

Accessory
Accessory

“Haiim” Vacuum-assisted blood collection system

Cassette (HC-001)

 

General Specifications

2. Cassette (HC-001)

Yellow light:

2 . E l e c t r i c a l   S a f e t y

2.1 Only use a power adapter having the model stated in Section E of this user 
manual. Any power adapter having a model different from the aforementioned 
model is prohibited.
2.2  Please  do  not  use  any  other  cables  or  accessories  not  approved  by  the 
manufacturer in this user manual to avoid negative influence on electromagnetic 
compatibility.
2.3 Do not connect this product to electric sockets with wet hands.
2.4 If abnormal behavior is observed due to EM disturbances, please relocate the 
main device accordingly.
2.5 There may be an interaction between the product and an electric equipment 
due to electromagnetic radiation. Itʼs recommended a safety distance of at least 
30  cm  between  the  product  and  electric  equipment,  especially  sensitive 
equipment.
2.6 This Product is medical electrical equipment that needs special precautions 
regarding  EMC  and  needs  to  be  installed  according  to  the  EMC  information 
provided.

-1-

-2-

1.1 Power Adapter Connector

1.2 Switch

1.3 Start-

Button 

1.5 Indicator Lights

1.4 Cassette

Connector

ON

OFF

ON

OFF

1.1. Power Adapter Connector: a connector shall be connected to a power adapter 
(GEM12I12-P1J) to have power input before blood collection.
1.2.  Switch:  the  switch  of  the  main  device.  By  moving  Switch,  the  main  device 
could be turned on or off.
        1.2.1. ON position.
        1.2.2. OFF position.
1.3. Start-Button: a button for starting a blood collection process by gently press. In 
case of emergency or accidents, press again the Start-Button for 1 second to stop 
the blood collection process.
1.4. Cassette Connector: a connector for connecting to the Main Device Connector.
1.5.  Indicator  Lights:  the  indicator  lights  consist  of  three  LED  lights,  showing 
different colors corresponding to different status as described below:

1. Main Device  (HD-001)

G.【Description of Key Items】

The function of each key item of the Product as labeled and briefly described 
below:

-3-

1.2.1

1.2.2

The  Indicator  Lights  show  the  white  color 
right after powering on the main device. Then, 
the  Indicator  Lights  will  turn  to  the  green 
color and go into standby mode.

Standby mode.
The Indicator Lights are green when the main 
device  is  in  a  normal  condition,  and  is  not 
collecting blood. 

The  yellow  color  means  negative  pressure 
does not reach the preset value.
The  Indicator  Lights  first  show  the  yellow 
color  once  pressing  the  Start-Button.  If  the 
pressure  does  not  reach  the  preset  value 
within  10  seconds,  the  Indicator  Lights  will 
turn to the green color, and the main device 
will go back to the standby mode. Please refer 
to Scenario 1 of Section L for troubleshooting.

Blood  collection  mode.  When  the  negative 
pressure reaches the preset value, the Indicator 
Lights  will  turn  from  the  yellow  color  to  the 
blue color and maintain at blue for 2 minutes to 
complete the blood collection process.
If the Indicator Lights turn from the blue color 
to the yellow color during the blood collection 
process, please refer to Scenario 2 of Section L 
for troubleshooting.

2.4 O-ring

2.5 Blood Collection Hole

A specific design

angle of the

Tube Connector

(6) Gloves

(7) Alcohol pad,

or other consuma-

bles for disinfection

(8) Cotton swab

(9) Cotton ball, or

other hemostatic

consumables

Storage

Conditions

Temperature  range:0~40℃

Humidity range:30~75% RH

Avoid direct exposure to sunlight

(5) Wrist pad

®

Отзывы: