User Manual - Rev AJ
California Instruments
Number
Message String
Cause
Remedy
executed
programming frequency when in
with mode of operation such as
DC mode.
-201
"Invalid while in local"
Command
ot in rem
state
before issuing GPIB commands.
issued but unit
Put instrument in remote state
is n
ote
-203
"Command protected"
Command
Some commands are supported
is locked out
by the unit but are locked out for
protection of settings and are
not user accessible.
-210
"Trigger error"
Problem wi
system.
th trigger
Unit could not generate trigger
for transient execution or
measurement.
-211
"Trigger ignored"
Trigger request has been
Trigger setup incorrect or unit
ignored.
was not armed when trigger was
received. Check transient
system or measurement trigger
system settings.
-213
"Init ignored"
Initialization
en ignore
request has
Unit was told to go to armed
be
d
state but was unable to do so.
Could be caused by incorrect
transient system or
measurement acquisition setup.
-220
"Parameter error"
Parameter not allowed.
Incorrect parameter or
parameter value. Check
programming manual for
allowable parameters
-221
"Setting conflict"
Requested
to program a DC offset while in
AC mode
setting conflicts
Check other settings. E.g. trying
with other setting in effect.
-222
"Data out of range"
Parameter
outside of
wable r
Check programming manual for
data
allo
ange.
allowable parameter values
-223
"Too much data"
More data received than
expected
Check programming manual for
number of parameters or data
block size
-224 "Illegal
parameter
value"
Pa
supported
rameter value is not
Check programming manual for
correct parameters
-226
"Lists not same length" One or mor
programme
length.
e transient lists
All lists must be of same length
d has different
or transient cannot be compiled
and executed.
-241
"Hardware missing"
N/A
N/A
-254
"Media full"
No storage
save settings or data.
make room.
space left to
Delete other settings or data to
-255
“Directory full”
To
dir
o many w
ectory en
from waveform memory to make
aveform
Delete one or more waveforms
tries
room.
-256
“File name not found”
Wa
dir
veform
ectory
requested not in
Check waveform directory for
waveform names present.
-257
“File name error”
Incorrect fil
definition.
ename
Too many or non ASCII
characters used in waveform file
-283
“Illegal variable name”
Variable name illegal.
Use ASCII characters only
-300
"Device specific error"
Hardware related error
Check hardware for proper
MX
Series
240
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...