User Manual - Rev AJ
California Instruments
SWEEP GROUP
If the SWEEP group is selected, the screen will appear as shown in Figure 9-24. The screen
has the following parameters that are unique to the group:
Figure 9-24: IEC 1000-4-13 SWEEP
is defined by
FRANGE. To change level, the test state must be off and user class must be
selection is required
idual range sweep
r
e
IN
p
leted, a display
of harmonics current versus frequency plot is available. To view the graph,
select the field and press the ENTER key.
HARMONICS GROUP
If the Harmonics group is selected, the screen will appear as shown in Figure 9-25. The screen
has the following parameters that are unique to the group:
1. LEVEL
Sets the percentage level of the interharmonics relative to the fundamental.
The level is fixed for the entire frequency range, which
selected.
2. FRANGE The range is selected by rotating the shuttle. The range
to change the level for the selected range. Also, an indiv
is possible by selecting the desired range and selecting RUN RANGE from
the RUN field.
3. STEP
The step size defines the interharmonics sweep points. Step size is fixed fo
the entire sweep range. The STEP size can be changed when the test stat
is off.
4. IHFREQ
This field will show the current interharmonic frequency when the test is
running. An interharmonic frequency can be selected using the shuttle. To
run a frequency without sweeping through the entire range, a RUN PO T
can be selected.
5. RESONANT This field is used to report the resonant points, if any, after running the swee
test. No editing is allowed in this field. When the test is comp
Figure 9-25: IEC 1000-4-13 Harmonics
1. LEVEL
Sets the percentage level of the harmonic relative to the fundamental. Each
level is associated with a harmonic number. To change levels, the test state
must be off and user class must be selected.
2. HARM NO Selecting a harmonic number using the shuttle knob will allow examination of
the LEVEL and PHASE of each harmonic and changing the level and phase
angle when USER class is selected.
3. PHASE
Sets the phase angle of the harmonics relative to the fundamental. To change
phase, the test state must be off and user class must be selected.
MX Series
217
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...