4.5.2 3Pi Controller Measurements
For MX Series with the -3Pi controller, the following four measurement screens are available:
Mode
AC
DC
AC+DC
MEASUREMENTS 1
VOLTAGE
AC rms voltage
DC Voltage
AC rms voltage
CURRENT
AC rms current
DC Current
AC rms current
FREQUENCY Frequency n/a
Frequency
POWER Real
power
n/a
n/a
MEASUREMENTS 2
VA POWER
Apparent power
power
Apparent power
PEAK CURR
Highest AC current
found
Highest DC current
found
Highest AC
current found
POWER FACT
Power factor
n/a
Power factor
CREST FACT
Crest factor
n/a
Crest factor
MEASUREMENTS 3
VOLT THD
Voltage distortion
Voltage distortion
n/a
CURR THD
Current
distortion
Current distortion
n/a
INST PK CURR
Instantaneous peak
Highest DC current
Instantaneous
t
current
found
peak curren
PHASE
Phase angle
n/a
Phase angle
The 3Pi controller has a fourth measurement screen for harmonics and trace analysis
measurements. This subject is covered in the next chapter.
Note: The V and I distortion calculations are based on H2 through H50 with the RMS current in
the denominator. Note that some definitions of THD use the fundamental component (H1) as the
es all the measurement parameters before
displaying the requested screen. This process may take up to a second. Consequently, pressing
the MEAS key may not always bring up the selected screen immediately. There will be a
perceptible delay. This will prevent the screen from appearing with invalid or blank readouts.
The measurement method for voltage and current will depend on the power source’s operating
mode. The following table shows the return value type (rms or average) and method of coupling
when the measurement command is initiated with a different extension at various operating
modes (AC, DC or AC + DC).
denominator. This may result in different readings between instruments depending on the
implementation chosen.
Measurements are always running in the background. When the user selects a measurement
screen for display, the AC source first updat
Operating Mode
Measurement
Extension
and Coupling
AC
DC
AC + DC
AC rms rms
rms
DC rms rms average
Coupling AC
DC
DC
MX Series
127
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...