User Manual - Rev AJ
California Instruments
UNDER FREQUENCY
This test will change the output frequency in the following sequence:
r 7 seconds.
•
380Hz for 7 seconds.
The
←
key (backspace) will terminate the test at any time.
9.6.1.8 DC
•
0Hz fo
TESTS
If the output voltage is set for 28V DC or 270V DC the MIL704 DC Main selection screen will
appear as seen in Figure 9-38.
Figure 9-38: MIL704 DC Menu
9.6.1.8.1 Steady State Test
Scroll to the STEADY STATE entry using the up and down cursor keys. Press the ENTER key
to select the STEADY STATE screen. The screen will appear as shown
Figure 9-39: Steady State DC
1 VOLTAGE
2 RIPPLE
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to
s
8V
system:
2. 270V
system:
•
250V for 1 minute.
•
280V for 1 minute.
•
270V for 1 minut
The MIL704 STEADY STATE screen has the following tests:
start the elected test.
VOLTAGE
This test will change the output voltage for the selected phase in the following sequence:
1. 2
•
22V for 1 minute.
•
29V for 1 minute.
•
28V for 1 minute.
e.
MX
Series
232
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...