5. Cut the green cable ties to the braided tubes that connect into the ATF pressure and vacuum ports as
shown below. Disconnect the braided tubes.
5a. Install port caps.
6. Unscrew the 3 screws (M4x16 Pan Head Machine Screw, P/N 6961-416) and remove the screws and
washers (M4 Flat washer, P/N 6985-04) as shown below.
Figure 37: Remove ATF Module from Eclipse Case.
7. Lift the ATF. Take off the 3 grommets. Disconnect the wire harness from the gearbox of the ATF.
8. Remove the ATF.
9. Install 3 Grommets into the mounting slots on the ATF as shown below. Position the ATF into the
bottom cover but DO NOT install the mounting screws yet. Route the 16” Silicone Tube under the ATF as
shown on Figure 39 on next page.
Figure 38: ATF Module Grommet Installation
Remove
Screws
Grommet
Caps
Caps