Personal Ambulatory Oxygen System
Provider Technical Manual
6
Oxygen Concentrator
Dimensions (H x W X D)
19.3 x 12.3 x 7.1 inches
(49.0cm x 31.2cm x 18.0cm)
Weight
Eclipse
Power Cartridge
15.0 pounds
3.4 pounds
Flow Settings
Continuous Flow
(measured in Liters Per Minute LPM)
Pulse Dose (measured in mL)
0.5 to 3.0 LPM (0.5 liter increments)
16-96 (16mL increments), 128mL, 160mL, 192mL
Continuous Flow Accuracy
+/- 10% or 200ml/min, whichever is greater
Oxygen Concentration
87–95.6% for all flow settings
Maximum System Pressure
15 psig (103.5 kPa)
Oxygen Output Pressure
5.0 psig (34.5 kPa) nominal
Oxygen Concentration
Status Indicator
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% ± 3%
Red Light Flashing = Abnormal Operation, less than 70% ± 5%
Nominal Sound Level
3.0 LPM Continuous Flow
3.0 Pulse Dose Mode
48 dB(A)
40 dB(A)
Operating Environment
Temperature
Humidity
50º F to 104º F (+10° to 40°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point
Storage Environment
Temperature
Humidity
-4º F to 140º F (-20° to 60°C)
Up to 95% Non-condensing
Altitude
0 – 13,123 feet (0 - 4,000 meters)
Nominal Power
0.5 LPM Continuous flow
3 LPM Continuous Flow
1.0 Pulse Dose Setting
6.0 Pulse Dose Setting
Battery Charging
48 Watts
145 Watts
45 Watts
95 Watts
50 Watts
Nominal Power Cartridge operating time
Continuous Flow
2.0 LPM
3.0 LPM
Pulse Dose (@ 12 BPM)
2.0
6.0
Continuous Flow Indication
Expressed in liters per minute (LPM)
Audible Alarm Indicators
• Loss of Power/Hot Power Cartridge
• Low Power Cartridge/Warm Power Cartridge
• Low Oxygen Output
• O
2
Flow Outside Normal Limits
• Unit Malfunction
• Low 9V battery
Back-Up Alarm Power
9V Internal Battery
Filters
Air Inlet, HEPA, Compressor Intake
Device Classification
IEC Class I, Type B Applied Part, IPX0
ECLIPSE OXYGEN SYSTEM SPECIFICATIONS
Note: Times will decrease with higher bolus size,
breath rate, ambient temperature, power cartridge
age and use over time.
2.0 hours
1.3 hours
5.2 hours
3.5 hours