ADJUSTING RISE TIME
The adjustable Rise Time feature on the Eclipse 3 was designed for patient comfort. The Rise Time feature adjusts flow and
speed of bolus delivery, and determines how quickly the patient receives their bolus volume while in Pulse Dose Mode. The
Eclipse 3 offers delivery settings of FAST, MEDIUM, and SLOW. The factory default setting is FAST, and is equivalent to the
delivery settings of the Eclipse and Eclipse 2 models. A MEDIUM or SLOW rise time may be appropriate for certain patients,
and should be determined and set by a trained clinician. Adjusting the Rise Time will not affect the chosen volume of oxygen
delivered to the patient (16-192 mL).
CONNECTING THE AC POWER SUPPLY
The Eclipse operates from external power when connected to a power outlet.
To connect the Eclipse to the AC Power Supply, follow these steps:
1. Insert the AC supply plug tip into the Eclipse.
2. Insert the power cord into the AC Power Supply.
3. Plug the power cord into a grounded outlet.
4. The power supply LED displays green to indicate that the AC Power Supply is drawing power. The External
Power Present Indicator on the Control Panel illuminates.
International safety standards require the use of hospital grade AC power cords (3893-SEQ) when using the Eclipse in a
medical care facility. Contact Chart SeQual to obtain a hospital grade AC power cord for the Eclipse if necessary.
NOTE:
If the Power Supply Status LED is not illuminated after inserting the plug
into a grounded outlet, check to make sure the Power Supply and power
cords are securely plugged into the Eclipse.
NOTE:
When removing the AC Power Supply from the Eclipse, remove the plug
from the AC outlet before removing the AC Power Supply plug from the
Eclipse.
CAUTION
• DO NOT connect the Eclipse to an extension cord or electrical outlet
controlled by a switch.
• Ensure adequate clearance around the AC Power Supply.
• The AC Power Supply is universal input, but the AC power cord
is appropriate to specific country’s electrical service. Ensure that
power cord is appropriate to country’s electrical service.
WARNING: