Appendix C

APPENDIX C - VARIABLES IN VISCOSITY MEASUREMENTS

As with any instrument measurement, there are variables that can affect a 

Viscometer measurement. These variables may be related to the instrument 

(Viscometer), or the test fluid. Variables related to the test fluid deal with the 

rheological properties of the fluid, while instrument variables would include the 

Viscometer design and the spindle geometry system utilized.

RHEOLOGICAL PROPERTIES

Fluids have different rheological characteristics that can be described by Viscometer 

measurements. We can then work with these fluids to suit our lab or process 

conditions.

There are two categories of fluids:

Newtonian

 

These fluids have the same viscosity at different Shear Rates 

(different RPMs) and are called Newtonian over the Shear Rate 

range they are measured.

Non-Newtonian

 These fluids have different viscosities at different shear rates 

(different RPMs). They fall into two groups:

 

1) Time Independent non-Newtonian

 

2) Time Dependent non-Newtonian

The time dependency is the time they are held at a given Shear Rate (RPM).  

They are non- Newtonian, and when you change the Viscometer spindle speed,  

you get a different viscosity.

Time Independent

Pseudoplastic

  A pseudoplastic material displays a decrease in viscosity with an 

increase in shear rate, and is also known as “shear thinning”. If 

you take Viscometer readings from a low to a high RPM and then 

back to the low RPM, and the readings fall upon themselves, the 

material is time independent pseudoplastic and shear thinning.

Time Dependent

Thixotropic

 

 A thixotropic material has decreasing viscosity under constant 

shear rate. If you set a Viscometer at a constant speed, recording 

P

 values over time, and find that the 

P

 values decrease with time, 

the material is thixotropic.

Содержание CAP 1000+

Страница 1: ...Additives Instruments A member of Measure what you see CAP 1000 Viscometer Manual...

Страница 2: ......

Страница 3: ...8 D 82538 Geretsried Germany Tel 0 800 gardner 0 800 4273637 49 8171 3493 0 Fax 49 8171 3493 140 BYK Gardner USA 9104 Guilford Road Columbia MD 21046 USA Phone 800 343 7721 301 483 6500 Fax 800 394 82...

Страница 4: ...Therefore in addition to our products we offer many VALUE ADDED services Technical Sales Force Technical Application Support Application and Technical Seminars Repair Certification Service BYK Gardner...

Страница 5: ...II 2 Cone Spindle Selection and Setting 14 II 3 Speed Setting 16 II 4 Temperature Control Setting 16 II 5 Hold Time Settings 17 II 6 Run Time 17 II 7 Printing 17 II 8 Run and Stop Keys 18 III OPERATIO...

Страница 6: ...6...

Страница 7: ...e control of sample is possible between either 5 C or 15 C below ambient whichever is higher and 75 C or 50 C and 235 C depending on viscometer model The CAP1000 can also be ordered with a single cust...

Страница 8: ...to be sure that you have received all components and that there is no damage If you are missing any parts please notify BYK Gardner or your local BYK Gardner representative immediately Any shipping da...

Страница 9: ...pindles and temperature plates are made of tungsten carbide Operating Environment CAP 1000 Viscometers must be operated within the following ambient temperatures 5 C 41 F to 40 C 149 F and humidity 20...

Страница 10: ...irements match your power source BEFORE connecting it to power The AC input voltage and frequency must be within the appropriate range as shown on the name plate of the viscometer NOTE The CAP Viscome...

Страница 11: ...the manual for specific warning or caution information to avoid personal injury or damage to the instrument SAFETY OVERVIEW If this instrument is used in a manner not specified by the manufacturer the...

Страница 12: ...Stops cone spindle rotation at any time Exits data entry field DELETE Overwrites entered data PRINT Sends data to the parallel port RUN Starts spindle rotation and measurement RUN TIME Selects time en...

Страница 13: ...ge condition If the viscometer settles out with a final reading below zero negative values will be displayed Temperature is displayed in C degrees centigrade I 8 Cleaning Instrument Keypad Painted Sur...

Страница 14: ...pressing the STOP key during instrument power up The CAP 1000 can be configured to operate at 750 RPM or 900 RPM This selection is set by choosing 1 SPEED in the special functions screen then selecti...

Страница 15: ...viscometer calculations The screen will display the following message Spindle 04 CALIBRATE YES NO Figure II 3 Normally there is no requirement to perform a cone calibration Cones supplied at the time...

Страница 16: ...The default temperature on start up is 25 0 C on low temperature models and 50 C on high temperature models The temperature ranges are Low temperature 5 C or 15 C below ambient whichever is higher to...

Страница 17: ...time of zero sets the viscometer to infinite run mode In this mode the spindle will rotate at the set speed for as long as the RUN button is pressed When the RUN key is released the spindle will stop...

Страница 18: ...imer setting 3 Used in executing a cone calibration The STOP key has three functions 1 Stops the cone rotation at any time 2 Pressing and holding the STOP ESCAPE key during power up selects the viscos...

Страница 19: ...er for accuracy information for CAP spindles 07 10 Table III 1 Accuracy of Viscosity Measurement CAP 1000 L VISCOMETERS Cone FULLSCALE VISCOSITY RANGE ACCURACY 750 RPM 900RPM Over 900 RPM to 1000 RPM...

Страница 20: ...viscosity measurement With the viscometer on a clean level surface connect it to the proper power supply Section I 4 1 Turn the power switch ON Section II 1 The procedure assumes that the following li...

Страница 21: ...e about 1 0 mm see table A 1 for sample size 3 Release the solvent trap placing it onto the plate over the cone so it does not touch the cone shaft The user should ensure that the substances placed un...

Страница 22: ...n the test conditions and viscosity results from the viscometer display 10 Relocate the solvent trap onto the cone adapter and raise the handle 11 It is recommended to remove the cone for cleaning How...

Страница 23: ...700 CAP7L 1328 cP CAP7H 1328 cP 08 2N 400 CAP8L 5313 cP CAP8H 5313 cP 09 2N 100 CAP9L 21 250 cP CAP9H 21 250 cP 10 5N 170 CAP10L 236 cP CAP10H 236 cP N RPM Approximate Value Table A 1 CAP 1000 VISCOSI...

Страница 24: ...viscosity range FSR for any cone speed divide the Cone Range Coefficient refer to Table A2 by the rotational speed RPM Full Scale Viscosity Range Poise Cone Range Coefficient Equation A1 RPM CAP 1000...

Страница 25: ...l scale viscosity range at 750 RPM Full Scale Viscosity Range poise 3 750 5 00 poise 750 Using Equation A5 determine the viscosity at 73 8 of full scale range Viscosity poise 5 00 poise x 73 8 3 69 po...

Страница 26: ...on INTERPRETATION OF CALIBRATION TEST RESULTS When verifying the calibration of the CAP 1000 the instrument and viscosity standard fluid error must be combined to calculate the total allowable error T...

Страница 27: ...solvent trap NOTE The solvent trap must be utilized when calibrating to ensure proper temperature control 7 Select the spindle to be calibrated by using the SPINDLE key 8 Select YES for CALIBRATE and...

Страница 28: ...cosity The SMC and SRC must be entered after entering the spindle number 11 20 and selecting YES 1 for calibration Follow the screen prompts Once SMC and SRC are entered follow the procedure described...

Страница 29: ...Rate range they are measured Non Newtonian These fluids have different viscosities at different shear rates different RPMs They fall into two groups 1 Time Independent non Newtonian 2 Time Dependent...

Страница 30: ...000 rpm the two viscosity values produced may be different because the readings were made at different shear rates The faster the spindle speed the higher the shear rate The shear rate of a given mea...

Страница 31: ...ipping damage The Viscometer should be shipped in its carrying case together with all spindles originally provided with the instrument For repair or service return to one of the locations listed below...

Страница 32: ...199 024 584 E 1610...

Отзывы: