15
Cone Spindle Selection and Setting
When using the solvent trap, connect it to the cone adapter by sliding it up,
passing the slot by the thumb screw and turning the trap clockwise onto the
thumbscrew
. Slide the cone up into the adapter as far as it will go and hand lock it
in place with the thumb screw. Tighten the thumb screw firmly and securely.
Press the
SPINDLE
key. The display will change to the spindle entry screen. Using
the number keys, type in the required spindle number.
Two digits must be entered for the cone number. For cone
01
through
09
, the first
number remains as “0”.
NOTE:
The default cone setting on power-up will be the last cone entry prior to
shutting off the viscometer.
After the correct two (2) digits have been entered, press the
ENTER
key and the
cone will be accepted for viscometer calculations. The screen will display the
following message:
Spindle 04
CALIBRATE? YES/NO
Figure II-3
Normally there is no requirement to perform a cone calibration. Cones supplied at
the time of order are calibrated to the viscometer prior to shipment.
NOTE:
1. CAP Viscometers allow for
only one
cone at a time of the
same cone
number
to be calibrated to the viscometer. Multiple cones of the same
cone number
must each be calibrated
to the viscometer before
operation (refer to
Appendix B
).
2. Cones entered as 11 through 99 must be first calibrated following the
directions in
Appendix B
.
If you are not going to calibrate the cone, continue by pressing the
NO
key then
ENTER
. The viscometer will display the
MAIN SCREEN (Figure II-1)
.
If you are going to calibrate the cone, press the
YES
key, refer to
Appendix B
, and
follow the calibration instructions under
Cone Calibration
.
Figure II-2
Thumb Screw
Solvent Trap
Cone Spindle
Viscometer in measurement position
Viscometer in „cleaning“ position