13
Viscosity and Temperature Display / Cleaning
I.7 VISCOSITY AND TEMPERATURE DISPLAY
Viscosity is displayed in either
P=Poise
or
cP=Centipoise
(CGS system) or
Pa•s=Pascal seconds
or
mPa•s=milliPascal seconds
(SI system). If the viscosity
measurement is over range, “
EEEE
” will be displayed. BYK-Gardner recommends a
minimum reading of 10%. If the display value is between 0 and 10%, the unit display
will flash to indicate an out of range condition. If the viscometer settles out with a
final reading below zero, negative values will be displayed.
Temperature is displayed in
°C=degrees centigrade
.
I.8 Cleaning
Instrument, Keypad & Painted Surfaces:
Clean with dry, non-abrasive cloth.
Do not use solventsor cleaners.
Immersed Components (spindles/cones) and insulation plate:
All immersed components are made of carbide steel.
Clean with non-abrasive cloth and solvent appropriate for sample material
that is not aggressive to immersed components.
NOTE:
When cleaning, take care not to apply excessive force which may
bend the spindle shaft or otherwise damage the instrument.