Accuracy of Measurement
III. OPERATION
III.1 ACCURACY OF MEASUREMENT
Table III.1 indicates the accuracy of the viscosity measurement with CAP spindles
1-6. This accuracy depends on both the rotational speed of the cone and the
percent of full scale range (in Poise) at which viscosity is measured. The accuracy
for viscosity data provided by CAP Viscometers is the indicated percentage of the
full scale range. See Appendix A for information on how to determine Full Scale
Viscosity Range (FSR).
Please contact BYK-Gardner for accuracy information for CAP spindles 07-10.
Table III-1 Accuracy of Viscosity Measurement
CAP 1000+L VISCOMETERS
Cone
FULLSCALE VISCOSITY RANGE ACCURACY
″750 RPM
″900RPM
Over 900 RPM to 1000 RPM
01
″ ± 2.0%
″ ±2.0%
Varies depending on the thermal conductivity of
the sample being measured
02
″ ± 2.0%″
″ ±2.0%
03
″ ± 2.0%
″ ±2.0%
04
″ ± 3.0%
″ ±4.0%
05
″ ± 4.0%
″ ±6.0%
06
″ ± 7.0%
″ ±10.0%
CAP 1000+H VISCOMETERS
Cone
FULLSCALE VISCOSITY RANGE ACCURACY
750 RPM
900RPM
Over 900 RPM to 1000 RPM
10 to 100%
FSR
″50% FSR
>50% FSR
FSR = Full Scale Range
01
″ ± 2.0%
″ ±2.0%
″ ±4.0%
Varies depending on the thermal
conductivity of the sample being
measured
02
″ ± 2.0%
″ ±2.0%
″ ±4.0%
03
″ ± 2.0%
″ ±2.0%
″ ±4.0%
04
″ ± 3.0%
″ ±3.0%
″ ±6.0%
05
″ ± 4.0%
″ ±4.0%
″ ±8.0%
06
″ ± 5.0%
″ ±5.0%
″ ±10.0%
Viscometer
Temperature Control Accuracy
CAP1000L & H
± 0.2°C