Type MS08
Maintenance
71
15.4.2
Cleaning the measuring window
Make sure that you do not touch the window surface with your fingers!
Clean the window with a few drops of acetone and a lint-free cloth, a clean paper towel or a special optical
paper.
ATTENTION!
▶
Do not use any aggressive cleaning solutions, putty, sandpaper or cleaning solutions that contain abras-
ive substances to remove dirt.
Damaged windows can be replaced by the Bürkert Customer Service. Please contact our Bürkert Customer
Service.
ATTENTION!
▶
After replacing the measuring windows, carry out a new zero-value measurement.
Fig. 25: Measuring window cleaning
15.4.3
Preparing the sensor for the function test and zero value
determination
Preparation
Preparation: Have a suitable measurement container ready nearby, cleaned carefully with a detergent solu-
tion and rinsed with ultra-pure water. The container must be filled with ultra-pure water so that the measur-
ing windows are completely covered by water.
Clean the probe as described in chapter
At the end of the cleaning process, rinse the probe carefully with deionized water.
Dry the sensor with a paper towel.
Remove any greasy residues: Wipe the sensor off with a little acetone on a kitchen towel.
CAUTION!
▶
For your own safety, you must wear the appropriate gloves and protective goggles!
Clean the sensor window according to the previous instructions in
Cleaning the measuring window [
Important: Polish the window next with a soft dry cloth or special optical paper to remove the thin film that
may have appeared while cleaning the window.