BCM-MS, BCM-LS
expected with gearboxes
and unfavourable install-
ation positions.
Measurements
outside the spe-
cification
Observe the technical
data and operate the
sensor within the spe-
cified measurement
ranges.
• No analog
output
Cable not prop-
erly connected
First check the electrical
connection of the sensor
and verify the data and
power cable are properly
connected. Please note the
prescribed pint assign-
ment.
Operating
voltage outside
the prescribed
range.
Please operate the sensor
within a range between
9 V and 33 V DC.
Interface config-
uration incorrect
Check and if necessary
correct the settings of the
analog outputs.
Analog outputs
wired incorrectly
Observe the information
on measuring the analog
outputs
• No sensor
communic-
ation via
CAN
Cable not prop-
erly connected
First check the electrical
connection of the sensor
and verify the data and
power cable are properly
connected. Please note the
prescribed pint assign-
ment.
Operating
voltage outside
the prescribed
range.
Please operate the sensor
within a range between
9 V and 33 V DC.
Interface config-
uration incorrect
Check and if necessary
correct the settings for the
interface parameters. The
correct setting depends
on the sensor configura-
tion.
CAN interface not
enabled
enable the CAN interface
with the RS232 port, CM-
SensorConfig or a ter-
minal program as de-
scribed in chapter Opera-
tion and Control.
• Absolute
humidity
measure-
ment incor-
rect
Calibrating para-
meter settings in-
correct
The calibrating paramet-
ers are specific to the oil
and must be pro-
grammed.
Please contact Bühler
Technologies GmbH Ser-
vice.
Measuring range
setting incorrect
The measuring range is
specific to the oil and
must be programmed.
Please contact Bühler
Technologies GmbH Ser-
vice.
Tab. 6:
Error: No sensor communication with HyperTerminal
14
Bühler Technologies GmbH
BX150104 ◦ 03/2021