Bühler technologies PKE 42 T Скачать руководство пользователя страница 9

Brief Instructions
PKE 42-52 ΔT

5 Operation and control

pin assignment

power supply S1

status output S2

S3

3 - GND

4 - 4-20 mA out

1

Fig. 2: 

A100049 Cooler electric supply

The clamping area has a diameter of 8-10 mm.

4.4 Signal outputs

The device provides different status signals. The max-
imum switching load of the alarm outputs is limited to
250 VAC/DC, 1 A each.

Alarm is set if the actual temperature of the cooler is
higher or lower than the preset alarm limits. The signal
does not distinguish if the alarm is caused by to high or
to low temperature.

When the moisture detector (optional) is installed, an
alarm is activated if the moisture is still present in the
prepared sample gas. Thereby, no distinction is made
between the alarm / cable break triggered by moisture
detector 1 or 2. This information is displayed by an error
message instead.

If the option "temperature signal" is installed, the actual
temperature of the connector is provided as an ana-
logue signal. (Option "moisture detector" includes the
option "temperature signal".) The signal is provided via
connector S2 (M12x1). This port is placed next to the
connectors of the moisture detector on top of the cooler.

5 Operation and control

NOTICE

The device must not be operated beyond its specifica-
tions.

After turning on the power supply the display shows the
actual temperature of the cooling block. The display
blinks until the (set) temperature range with respect to
the pre-set output dew point is reached. The status con-
tact is switched to "Alarm".

If the temperature range is reached, the actual temper-
ature is shown constantly and status contact switches
back.

The same applies to operation mode ∆T-mode. If the
pre-set temperature difference is reached, the actual
temperature is permanently displayed and the status
contact is reset.

If the display starts blinking during operation or if an er-
ror message is displayed see chapter "Troubleshoot-
ing".

For performance limits see datasheet.

5.1 Operation of the menu functions

Overview of the operational principal:

Use this short description if you have experience with
the device.

Operation is carried out by only the keys with the follow-
ing functions:

Key

Function

• Switch from measurement display to main

menu

• Selection of the display menu item
• Accepting the changed value or selection

• Switch to the upper menu item
• Increase of the value of switching the se-

lection

• Temporary display of the alternative meas-

urement display (if option is installed)

• Switch to lower menu item
• Decrease of the value of switching the se-

lection

• Temporary display of the alternative meas-

urement display (if option is installed)

5.1.1 Overview of the menu items

9

BX440013

Содержание PKE 42 T

Страница 1: ...rzanleitung Peltier Messgask hler deutsch 2 Brief Instructions Peltier sample gas cooler english 7 Brief Instructions Refroidisseur de gaz de mesure Peltier fran ais 11 Brief Instructions Refrigerador...

Страница 2: ...er meiden Sie dadurch Personen und Sachsch den Der Betreiber der Anlage muss sicherstellen dass Sicherheitshinweise und Betriebsanleitungen ver f gbar sind und eingehalten werden die Pr fungen vor Inb...

Страница 3: ...st der Kondensatableiter mittels einer Schelle separat zu befestigen Der Kondensatableiter kann direkt am W rmetauscher befestigt werden Kondensatleitungen sind grunds tzlich mit Gef lle und Mindestne...

Страница 4: ...gel st wenn im aufbereiteten Messgas noch Feuch tigkeit enthalten ist oder ein Kabelbruch erkannt wird Dabei wird nicht unterschieden ob der Alarm Kabel bruch durch den Feuchtef hler 1 oder 2 ausgel s...

Страница 5: ...ahrenheit dargestellt werden Anzeige der Temperatur und des Betriebszustandes ____ ____ ____ Hauptmen Untermen Einstellbereich speichert den Wert ____ Anzeige ____ aktuelle Temperatur ____ Wahlmen ___...

Страница 6: ...eise zur Wartung finden Sie in der Originalbetriebsanleitung auf der beigef gten CD oder im Internet unter www buehler technolo gies com 7 Service und Reparatur Eine ausf hrliche Beschreibung des Ger...

Страница 7: ...e against damage The operator of the system must secure that safety and operation instructions are accessible and followed local safety regulations and standards are obeyed performance data and instal...

Страница 8: ...d at the same time are already installed and connected to the peristaltic pumps NOTICE Installing peristaltic pumps limits the maximum ap proved operating pressure in the system Operating pressure 0 5...

Страница 9: ...beyond its specifica tions After turning on the power supply the display shows the actual temperature of the cooling block The display blinks until the set temperature range with respect to the pre s...

Страница 10: ...djust the set point 2 C 20 C 35 6 F 68 F Set the upper alarm threshold above setpoint tEMP 1 C 7 C 1 8 F 12 6 F Set the lower alarm threshold below setpoint tEMP 1 C 3 C 1 8 5 4F If moisture detector...

Страница 11: ...venez les pannes et vitez ainsi des dommages sur du mat riel et des personnes L exploitant du dispositif doit s assurer que les indications de s curit et le mode d emploi soient disponibles et respec...

Страница 12: ...geur de condensat direc tement sur le tube de raccordement Pour les tuyaux il est n cessaire d accrocher le purgeur de condensat s par ment l aide d un collier de serrage Le purgeur de condensat peut...

Страница 13: ...ble est d tect e L alarme la rupture de c ble est d clench e de mani re indiff renci e par les capteurs d humidit 1 ou 2 Ce pendant cette information appara t sur l affichage Si l option signal de te...

Страница 14: ...st r se d alarme Etalonner le capteur d humidit Quitter le sous menu C F R glages globaux Quitter le menu principal R glage de temp rature de refroidisseur dans la plage 2 C 20 C 35 6 F 68 F Placer la...

Страница 15: ...normativas de seguridad relevantes en el lugar de montaje as como las regula ciones generales de la t cnica Prevenga las aver as evitando de esta forma da os personales y materiales El usuario de la i...

Страница 16: ...terial ser necesario establecer una conexi n mediante uniones roscadas y un tubo o man guera entre el intercambiador de calor y el purgador de condensados Si tratamos con acero el purgador de condensa...

Страница 17: ...do todav a contenga humedad o en caso de detectar la rotura de alg n cable En estas si tuaciones no se diferenciar si la alarma rotura del ca ble ha sido reconocida por el sensor de humedad 1 o 2 Toda...

Страница 18: ...global Abandonar el men principal Ajuste de la temperatura del refrigerador en valores 2 C 20 C 35 6 F 68 F Establecer el l mite de alarma superior en valores tEMP 1 C 7 C 1 8 F 12 6 F Establecer el l...

Страница 19: ...ler technologies com Harkortstra e 29 40880 Ratingen Deutschland 49 0 2102 4989 0 49 0 21 02 49 89 20 1 1 1 2 1 3 CD 2 BetrSichV GUV V A1 GUV V A3 a b c d a b c 3 20 C 60 C 4 4 1 10 cm 4 2 IN 11 LD sp...

Страница 20: ...E 42 52 T 5 DN8 10 5 16 0 5 bar 4 2 1 G DN 4 6 1 4 1 6 AGF PV 30 F2 G 2 bar 4 2 2 4 2 3 PTFE PTFE 4 3 DIN43650 S1 S2 S3 3 GND 4 4 20 mA out 1 2 A100049 8 10 mm 4 4 250 VAC DC 1 A 1 2 M12x1 5 20 BX4400...

Страница 21: ...__ ____ ____ ____ ____ ____ H2O ____ ____ ____ P52_ ____ ____ ModE ____ ____ 5s C F 2 C 20 C 35 6 F 68 F tEMP 1 C 7 C 1 8 F 12 6 F tEMP 1 C 3 C 1 8 F 5 4 F T tEMP A Hi A Lo h2o ModE E rSt ___ ___ ____...

Страница 22: ...PKE 42 52 T 6 6 CD www buehler technologies com 7 CD www buehler technologies com 22 BX440013...

Отзывы: