21
BÜCHI
Rotavapor
R-187EL
6 Maintenance
All regulations aimed at keeping the Rotavapor in good working
order must be observed.
They include cleaning the appliance at regular intervals and
examining it for signs of damage.
Before carrying out any maintenance work on the unit, make sure
that the steam and electricity supplies are disconnected. For re-
pairs to the bottom side of the bath, always make sure the bath
is well supported or dismantle the bath pan.
6.1 Changing the hydraulic oil
Change the hydraulic oil at intervals of approx. 3000 operating
hours.
Hose
20 litre vessel
Pump
The oil must be fully emptied to carry out any repairs to the
hydraulic unit.
Possible hydraulic oil brands
Agip OSO 45
NUTO H 46
Aral Vitam HF 46
FINA Hydran 46
Aral Vitam DE 46
RENOLIN MR 15
AVILUB Hydr.-Oil RSL 46
RENOLIN B 15
AVILUB Hydr.-Oil HVI 46
GULF Harmony 46AW
BLASOL 158
HYDROLUBRIC VG 46
BP Energol HLP 46
Mobil D.T.E. 25
BP Ernergol HLP-D 46
OEST Hydraulikoil H-LP 46
CASTROL HYSPIN AWS 46
OEST Hydraulikoil 46 DD
Chevron EP Hydraulic Oil 46
SHELL Tellus Oil 46
ELF Acantis 46
SUNVIS 821-WR
Hydrelf 46
Rando Oil HD B-46
Table 4: Table of hydraulic oils
Pumpe /
pump
Figure 20: Oil change device
6 Maintenance