Operating procedure
m
μ
m
μ
m
μ
2
3
4
5
1
6
[CURVE/FILTER/
TOL/CUST]
[n/ENT]
[PARAMETER]
from
step 9
[n/ENT]
[CUTOFF
↑
],
[
↓
]
or
[REMOTE
←
]
Ra
m
μ
m
μ
In the measurement mode press the [CURVE/FILTER/TOL/
CUST] key three times.
The GO/NG judgment function setup mode is entered.
Press the [n/ENT] key.
If this function has already been set, then the set value will
be displayed.
Press the [PARAMETER] key, if required, to switch to the
parameter for which the GO/NG judgment function is to be set.
If there is no need to switch from the current parameter, proceed
to the next step without pressing the [PARAMETER] key.
The currently flashing parameter changes.
While the desired parameter is flashing, press the [n/ENT]
key.
The objective parameter of the GO/NG judgment is accepted.
Enter the upper limit.
TIP
For information on how to enter a numeric
value refer to “2.2 Key Operations to Enter
Numeric Values”.
During numeric entry if the [MODE/ESC] key is
pressed before the [n/ENT] key has been
pressed, the entered value is canceled and the
measurement mode is restored.