Title: New Brunswick BioFlo 3000 Bioreactor SOP

 

 

8.12.29.

 

Wait for the pH value to reach its setpoint before proceeding. 

8.13.

 

Activate additional feed loops at the appropriate time as indicated by the process 
SOP.

   

8.13.1.

 

Activate a feed loop by selecting the box in the “Loop Name” row of the “pH 1” 
column. This box should be displaying the column heading “pH 1”.  Press the 
ALTER key until the “Feed 2” column heading appears (or the appropriate feed 
column heading if not “Feed 2”). 

8.13.2.

 

Enter the appropriate setpoint for the feed solution, as indicated by the process 
SOP, by choosing the box in the “Set” row of the “Feed 2” column.  Use the 
number keypad to enter the appropriate setpoint value and then press the ENTER 
key.   

8.13.3.

 

In the “Control” row of the “Feed 2” column press the ALTER key until “on” is 
displayed, then press ENTER. 

8.14.

 

Calibrate the Dissolved Oxygen Probe

 

Note:  The DO probe should not be calibrated until the desired working temperature has been 
reached.  

8.14.1.

 

Polarize the DO probe for 30 minutes prior to performing calibration.  

8.14.1.1.

 

If necessary, connect the DO cable to the DO electrical outlet on the front of 
the cabinet. 

8.14.1.2.

 

Remove the protective cap from the top of the DO probe and connect the 
DO Cable to the DO probe. 

8.14.1.3.

 

If necessary, turn on the BioFlo 3000. 

8.14.1.4.

 

Allow the probe to sit for at least 30 minutes. 

8.14.1.5.

 

Verify that the DO level is above 0. If the DO display reads 0 or below there   
may be a problem with the DO probe. Contact the lab manager if this 
occurs.  

8.14.2.

 

Use the SCREEN key to choose the <Calibration> screen and then press the 
ENTER key. 

8.14.3.

 

Use the arrow keys to choose the box in the “DO” row of the “Function” column. 
This box should be displaying “Read”. 

8.14.4.

 

Change “Read” to “Zero” by pressing the ALTER key until “Zero” appears and 
then pressing the ENTER key. 

8.14.5.

 

Use the arrow keys to choose the box in the “DO” row of the “Zero” column.  
Enter “0.0” using the number keypad and then press the ENTER key. 

8.14.6.

 

Use the SCREEN key to choose the <Master> screen and then press the ENTER 
key. 

8.14.7.

 

Use the arrow keys to choose the box in the “Set” row of the “Agit 1” column. 

8.14.8.

 

Enter “1000” (for 1000rpm) using the number keypad and then press the ENTER 
key. This will turn on the agitation motor. 

8.14.9.

 

Press the SCREEN key and then use the arrow keys to select the <Gases> screen 
and then press the ENTER key. 

8.14.10.

 

Set the gases mode to “Manual” by selecting the box in the “Mode” row (it is the 
only box available in the “Mode” row).  Press the ALTER key until “Manual” is 
displayed, then press the ENTER key. 

18

Содержание BIOFLO 3000

Страница 1: ...llow the SOP as described and to inform the instructor about any deviations or problems that may occur while performing the procedure 4 References 4 1 New Brunswick BioFlo 3000 Bioreactor instrumentat...

Страница 2: ...the DO pH and temperature cables to their appropriate outlets on the front side of the cabinet of the BioFlo 3000 8 2 2 If necessary insert the plastic plugs into the holes in the glass vessel 8 2 2 1...

Страница 3: ...sion with the harvest tube 8 3 8 Secure the harvest tube in place with a nut by lowering the nut over the top of the harvest tube and hand tightening it to the threads on the topside of the headplate...

Страница 4: ...the water The water valve is located under the lab bench where the BioFlo 3000 is located The pressure gauge should read between 15 and 20psi when the water is turned on 8 8 Calibrate the pH Probe 8 8...

Страница 5: ...O electrolyte in the bottom housing 8 9 5 Replace the bottom housing 8 9 6 Carefully insert the DO probe into the DO port of the headplate Tighten the adaptor with a wrench 8 9 7 Ensure that the DO pr...

Страница 6: ...the lab bench The pressure gauge should read between 15 and 20psi 8 12 9 Add approximately 1mL of glycerol to the thermowell and insert the RTD temperature probe into the thermowell 8 12 10 Remove the...

Страница 7: ...rection Therefore the solution will flow from right to left in the tube 8 12 24 Repeat step 8 12 23 for additional feed solution flasks if needed 8 12 25 Set the pH setpoint to the desired value Refer...

Страница 8: ...tlet on the front of the cabinet 8 14 1 2 Remove the protective cap from the top of the DO probe and connect the DO Cable to the DO probe 8 14 1 3 If necessary turn on the BioFlo 3000 8 14 1 4 Allow t...

Страница 9: ...e Agit 1 column Press the ALTER key until P I D appears and then press the ENTER key 8 15 1 4 To set the minimum rpm select the box in the Set row of the Agit 1 column Use the number keypad to enter t...

Страница 10: ...set the minimum rpm select the box in the Set row of the Agit 1 column Use the number keypad to enter the desired minimum rpm as indicated by the process SOP and then press the ENTER key 8 15 3 5 Set...

Страница 11: ...then press the ENTER key 8 16 8 Use the arrow keys to select the box in the Control row of the Feed 2 column 8 16 9 Set the control mode to Level 1 by pressing the ALTER key until Lvl 1 appears and t...

Страница 12: ...e 21Feb08HSA 8 17 6 7 Skip step 8 17 7 and continue with the SOP at step 8 17 8 8 17 7 To create a new recipe from the BioCommand Lite window select Create a new recipe with Batch Wizard and click Co...

Страница 13: ...ck OK 8 17 7 19 8 Click 8 17 7 20 Repeat step 8 17 26 until all parameters needed are added to the Recipe B1 Trend window AGIT1 DO2 1 TEMP 1 and pH 1 should always be added 8 17 7 21 Set the graph ran...

Страница 14: ...ean with a soft brush if needed Rinse thoroughly with tap water and then with deionized water 8 22 8 Remove the vessel from the heat exchanger base and pour the bleach solution down the drain Rinse th...

Страница 15: ...oughly 8 22 23 Clean the heat exchanger base and rinse thoroughly 8 22 24 Ensure that all vessel parts have been cleaned and rinsed thoroughly Allow them to air dry 8 22 25 Wipe down the BioFlo cabine...

Страница 16: ...nger Connectors 4 Bottom Steel Ring 5 Glass Vessel 6 Condenser Water Lines 7 Top Steel Ring 8 Agitator Shaft 9 Agitator Cover 10 Condenser Top Outlet 11 Electrical Outlet 12 Sparger Gas Port 13 Agitat...

Страница 17: ...09 Page 17 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 2 Vessel Parts 1 Small Thumb Screw 2 Large Thumb Screw 3 Condenser 4 Agitator Cover 5 Agitator 6 Headplate 7 Sampler Assembly Bul...

Страница 18: ...18 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 3 Headplate Underside 1 Addition Port 2 Thermowell Port 3 O ring 4 Foam Sensor Port 5 pH Probe Port 6 Agitator Shaft 7 Upper Impeller 8 L...

Страница 19: ...k BioFlo 3000 Bioreactor SOP Figure 4 Assembled Headplate Top Side 1 DO Probe 2 pH Probe 3 Foam Sensor behind pH Probe 4 Condenser 5 Condenser Barbed Connectors 6 Ground Stud 7 Thermowell 8 Addition P...

Страница 20: ...ve Date 20May09 Page 20 of 20 Title New Brunswick BioFlo 3000 Bioreactor SOP Figure 5 BioFlo 3000 Side View 1 Gas and Air Console 2 Cabinet 3 Power Switch 4 Feed Pump Figure 6 BioFlo 3000 Controller 1...

Отзывы: