Planning Details
60
Site Planning Information BioSpec 70/30 USR_2_V020
Figure 6.14: TN-S connection diagram of the MR instrument (60Hz)
1a
Earthing point of the building to be connected to the Bruker Line Power Distributor
(2) directly, or via an additional equipotential bonding bar (1b).
1b
Additional equipotential bonding bar to connect additional equipment other than
MR instrument components.
2
Bruker Line Power Distributor providing outgoing line power to all MR instrument
consumer components and equipotential bonding.
3
Power connection
4, 5
Generator or transformer of the building to provide the mains voltage required.
These units must be provided by the System Owner (responsible body).