Bronkhorst®
Instruction Manual Vapor Delivery Module
9.17.079D
18
4.6.2.3 Stopping process
Procedure
Actions
C1
Disconnect process
·
Open V6, close V7
C2
Stop liquid flow
·
Give setpoint = 0 to LFC
C3
Optional: Stop gas flow
·
Give setpoint = 0 to MFC
Make sure that the carrier gas is always flowing when
the liquid flow is on. Preferably, let always gas flow in
the carrier gas line when the system is filled with liquid.
4.6.2.4 Leaving system overnight
Procedure
Actions
D1
Relieve the supply pressure from the liquid- and
carrier gas lines
·
Close V2
·
Open V8 for some seconds
4.6.2.5 Emptying and purging system
Procedure
(1,2)
Actions
E1
Relieve the supply pressure from the liquid- and
carrier gas lines
·
Close V2 and V3
E2
Purge liquid out of lines
·
Open V1 and V4
E3
Purge lines with inert gas
·
Open V4 and V6
·
Close V7
·
Give setpoint to MFC
E4
Evacuate liquid lines (cont.)
·
Close V1
·
Open V8
E5
Optional: repeat last two steps for a few times
(1)
If the liquid is reactive or aggressive, be sure to remove air (and water vapor) carefully, before introducing liquid in the
system.
(2)
If the liquid is reactive, aggressive, poisonous, toxic, or flammable, beware that the system is purged fully (preferably with
inert liquid, E3, E4, E5) before opening the system.
·
In case of non-pure liquids, place a filter upstream the liquid flow controller.
·
Mount the VDM upright.