Bronkhorst®
Instruction Manual ES-FLOW™
9.17.145E
22
*) The following control modes are available:
Value
Description
RS-232
Normal operation via RS-232
bus/rs232
Normal operation via fieldbus or RS-232
analog input
Normal analog operation
setpoint 0 perc.
Setpoint set to 0%
setpoint 100 perc.
Setpoint set to 100%
control idle
Controller disabled, valve frozen in current position
actuator 0 perc.
Controller disabled, valve closed
actuator 100 perc.
Controller disabled, valve fully open
actuator steering
Controller disabled, valve opening equal to setpoint (percentage)
fb slave
Acting as slave of other instrument on FLOW-BUS
analog slave
Acting as slave of other instrument on analog input
fb ana slave
Acting as slave of other instrument on FLOW-BUS, slave factor set by analog input
signal
**) The
Master slave
item is only available if one of the slave control modes is selected. Availability of
Master node
and
Slave
factor
depends on the selected mode:
Control mode
Master node
Slave factor
fb slave
analog slave
fb ana slave
See also
Master/slave configuration (FLOW-BUS)
for more information about setting up a master/slave relationship between
instruments.
3.3.3.3
Counter/totalizer
Bronkhorst® flow meters have a built-in counter function, which can be used to monitor and/or control the amount of
media flowing through the instrument. The flow can be stopped or changed when a certain limit is reached. Until the
counter is reset, the counter setpoint overrides the regular setpoint.
·
Some menu items are not available on ES-FLOW models before Mk II.
·
On models before Mk II, the names of some menu items differ slightly from the descriptions below, but they are still easily
recognizable.