Bronkhorst®

Instruction Manual Controlled Evaporator and Mixer

9.17.126A

11

2

Installation

2.1

Functional properties

Before installing the CEM, check the serial number label to see if the functional properties match

your requirements:

·

Instrument type (technical properties; see 

Model key

)

·

Flow rates (gas and liquid)

·

Media to be used in the instrument

·

Power supply

Test pressure

Bronkhorst® instruments are pressure tested to at least 1.5 times the specified working process
conditions and outboard leak tested to at least 2 * 10

-9

 mbar l/s Helium. The tested pressure is specified

on the instrument with a red label.

·

Before installation, make sure that the tested pressure is in accordance with the safety factor of your application

·

If the pressure test label is missing or if the specified pressure is insufficient, the instrument must not be used and should

be returned to the factory

·

The maximum operating pressure must never exceed the tested pressure

·

Disassembling the instrument and/or replacing parts of it will invalidate the pressure test specification.

2.2

Mounting

The CEM must be mounted vertically, with the control valve up and the vapour outlet pointing down (as shown in the 

product overview

).

For optimal performance, please observe the following:

·

Use the mounting holes on the rear to attach the CEM to a rigid and stable construction. Check the 

dimensional drawing

for the exact positions and size of the mounting holes

·

Avoid installation in close proximity of mechanical vibration and/or heat sources.

2.3

Electrical connection

Electrical connections must be made with standard cables or according to the applicable hook-up diagrams. Make sure that

the power supply is suitable for the power ratings as indicated on the serial number label or in the technical specifications

of the instrument, and that double or reinforced insulation is used for the power supply.

The device contains electronic components that are susceptible to damage by

 electrostatic discharge

. Proper handling

procedures must be followed during installation, (dis)connecting and removing the electronics.

The device described in this manual carries the CE-mark and is 

compliant with the concerning EMC requirements

.

However, compliance with the EMC requirements is not possible without the use of proper cables and connector/gland

assemblies. Bronkhorst recommends the use of their standard cables. These cables have the right connectors and if loose

ends are used, these are marked to help prevent wrong connection. When using other cables, cable wire diameters should

be sufficient to carry the supply current, and voltage loss must be kept as low as possible. When in doubt, contact your local

Bronkhorst representative.

When connecting the product to other devices, be sure that the integrity of the shielding is not affected; 

always use

shielded cabling for signals and communication and do not use unshielded wire terminals.

Содержание CEM W101A

Страница 1: ...TENTION Pleaseread this Instruction Manualcarefully beforeinstalling and operating theinstrument Notfollowing theguidelines could resultin personalinjury and ordamageto theequipment Controlled Evaporator and Mixer CEM Instruction Manual ...

Страница 2: ...elpful information Thisinformationwill facilitatetheuseoftheinstrument and or contributetoitsoptimal performance Additional infoavailableontheinternet or from your local Bronkhorst representative Receipt of equipment Check the outside package box for damage incurred during shipment If the box is damaged then the local carrier must be notified at once regarding his liability if so required At the s...

Страница 3: ...s intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating information carefully before using the product Before operating make sure the line cord is connected to a properly grounded power receptacle Inspect the connecting cables for cracks or breaks before each use The equipment and access...

Страница 4: ...Bronkhorst Instruction Manual Controlled Evaporator and Mixer 9 17 126A 4 ...

Страница 5: ...1 12 Basic assembly 2 4 2 13 Working under vacuum 2 4 3 13 2 5 Fluid supply 13 Carrier gas 2 5 1 13 Liquid purity 2 5 2 13 Minimizing gas dissolution 2 5 3 14 Liquid vessel size 2 5 4 14 2 6 Media compatibility 15 Operation 3 15 3 1 Powering up 15 3 2 Required setpoints 16 3 3 Starting vapour production 16 3 4 Stopping vapour production 16 3 5 Bleeding mixing valve 17 3 6 Purging 18 3 7 Digital pa...

Страница 6: ...Bronkhorst Instruction Manual Controlled Evaporator and Mixer 9 17 126A 6 ...

Страница 7: ...sasdescribedintheMaterial SafetyDataSheetsofthemediatobeusedinthesystem ifapplicable ThewettedmaterialsincorporatedintheCEMarecompatiblewithmediaandconditions e g pressure temperature as specifiedat orderingtime Ifyouareplanningtousetheproduct includinganythirdpartycomponentssuppliedby Bronkhorst suchaspumpsor valves withother mediaand or other conditions alwayscheckthewettedmaterials includingsea...

Страница 8: ...atomizes the liquid and adds it to the carrier gas creating an aerosol which is heated by the CEM until it transforms into a vapour To monitor the internal temperature of the heat exchanger the CEM incorporates a PT100 temperature sensor An internal safety switch prevents overheating of the heat exchanger by interrupting the control signal as soon as the temperature reaches 200 C To control the li...

Страница 9: ...wingtableisavailableontheCEM product pagesunder www bronkhorst com products Type Document name Document no Brochures CEM Brochure 9 60 038 Manuals Instruction Manual Controlled Evaporator and Mixer this document 9 17 126 Instruction manual E 8000 9 17 076 Hook up diagrams Hook up diagram CEM W101A W102A W202A 9 16 086 Hook up diagram CEM W303B 120V 9 16 116 Hook up diagram CEM W303B 230V 9 16 115 ...

Страница 10: ...orator and Mixer 9 17 126A 10 1 6 Model key The model key on the serial number label contains information about the technical properties of the instrument as ordered The actual properties of your instrument can be retrieved with the diagram below ...

Страница 11: ...kthedimensional drawing for theexact positionsandsizeofthemountingholes Avoidinstallationincloseproximityofmechanical vibrationand or heat sources 2 3 Electrical connection Electrical connections must be made with standard cables or according to the applicable hook up diagrams Make sure that the power supply is suitable for the power ratings as indicated on the serial number label or in the techni...

Страница 12: ...erature pressureandvibrations Tubing diameterand flexibility Usetubingwithaminimal diameter that still supportstherequiredmediaflow thiswill helpkeepthefillingand refreshingtimeofthefluidic linesasshort aspossible Preferablyusehardtubing asopposedtoflexible toprevent internal volumechangesandreducetheimpact of ambient pressurefluctuations Insulation and heat tracing Applyinginsulationmaterial onth...

Страница 13: ... your process media cannot be ruled out installing a particle filter in the carrier gas supply line upstream from the flow controller is strongly recommended If necessary install a de humidifier in the carrier gas supply line upstream from eventual filters Humidity in the carrier gas can cause chemical reactions in the mixing valve in combination with certain liquids Purging the fluidic system bef...

Страница 14: ...d vessel size The liquid vessel should be large enough to provide a stable flow for a sufficient amount of time between refills Purging or flushing the fluidic system can consume a relatively large amount of liquid take this into account when selecting a suitable vessel size The table below gives an indication of the liquid consumption based on different flow rates Flow rate Liquid consumption g h...

Страница 15: ...c system graduallyuptothelevel of operating conditions openandclosethefluidsupplygently 3 2 Required setpoints Therequiredsetpointsfor thecarrier gasandliquidflowsandthetemperatureoftheheat exchanger canbeobtainedfrom FLUIDAT ontheNet usingtheflowratesoftheorderedinstruments FLUIDAT ontheNet canbeaccessedviahttp www fluidat com FreeregistrationontheFLUIDAT websitegives accesstoall availablefluidsa...

Страница 16: ...point to 0 C 3 Change the carrier gas flow setpoint to 0 3 5 Bleeding mixing valve Inorder toensurestableandreliablecontrol behaviour theliquidtubinganddeadspaceofthemixingvalveshouldbe kept freeofgas bybleedingthemixingvalvefrom timetotime Bleedingisrecommendedat thefollowingoccasions o prior tofirst useoftheCEMsystem o whenrestartingthesystem after purging o periodically toremoveaccumulatedgas i...

Страница 17: ...atility etc longer or shorter timesmight berecommendable Step 1 Preparation a Set all setpoints to 0 liquid flow carrier gas flow CEM temperature b Connect a purging gas supply facility to the inlets of both flow controllers liquid and gas c Pressurise the purging gas supply line Step 2 Gastubing a Change the carrier gas flow setpoint to 100 b Purge for at least 30 minutes Step 3 Mixing valve a Ch...

Страница 18: ...re than the actual heater temperature effectively switches the heater off Fmeasure Type Access Range FlowDDE FLOW BUS Modbus Float R 0 200 205 33 0 0xA100 0xA101 41217 41218 Fmeasure returns the measured temperature in C Fsetpoint Type Access Range FlowDDE FLOW BUS Modbus Float RW 0 200 206 33 3 0xA119 0xA11A 41241 41242 Fsetpoint sets the required temperature of the heater in C Note that the heat...

Страница 19: ...d with a clean dry and inert gas In case of severe contamination cleaning the inside of the device may be necessary by flushing the conduits with an appropriate cleaning fluid Inexpertlyservicinginstrumentscanleadtoseriouspersonal injuryand or damagetotheinstrument or thesystem it is usedin Therefore servicingmust beperformedbytrainedandqualifiedpersonnel Contact your local Bronkhorst representati...

Страница 20: ...gas flow too high Decrease carrier gas flow Liquid flow too low Increase liquid flow Condensation at vapour outlet CEM temperature setpoint too low Increase CEM temperature setpoint Carrier gas flow too low Increase carrier gas flow Liquid flow too high Decrease liquid flow CEM heater does not warm up Fuse blown Replace fuse Temperature sensor defective Return equipment to factory Heater defective...

Страница 21: ...he right product for your application Contact sales by e mail sales bronkhorst com For after sales questions our Customer Service Department is available with help and guidance To contact CSD by e mail support bronkhorst com No matter the time zone our experts within the Support Group are available to answer your request immediately or ensure appropriate further action Our experts can be reached a...

Страница 22: ...nloadedfrom theService Support sectionoftheBronkhorst website www bronkhorst com Important Clearly note on top of the package the customs clearance number of Bronkhorst High Tech B V NL801989978B01 only if applicable otherwise contact your local Bronkhorst representative for local arrangements 6 2 Disposal end of lifetime Within the scope of the European Union manufacturers of electrical and elect...

Отзывы: