Instruction Manual Controlled Evaporator and Mixer
9.17.126A
18
3.7
Digital parameters
Using an E-8000 CEM control module, the heating element of the CEM can be operated digitally via RS232 or fieldbus. This
section describes the parameters that are used to monitor and control the heating element.
Consult the
E-8000 manual
for installation information and a description of the user interface.
Measure
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
R
0…32000
8
1/0
0x0020/33
This parameter indicates the temperature measured by the CEM. The value range corresponds to 0…200 °C (the supported
temperature range of the CEM control module).
Setpoint
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
RW
0…32000
9
1/1
0x0021/34
This parameter sets the required temperature of the heater. Like
Measure
, its value range corresponds to 0…200 °C. Note that
the heater has no cooling functionality; a setpoint evaluating to a lower temperature than the actual heater temperature
effectively switches the heater off.
Fmeasure
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
R
0…200
205
33/0
0xA100…0xA101/ 41217…41218
Fmeasure
returns the measured temperature in °C.
Fsetpoint
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…200
206
33/3
0xA119…0xA11A/ 41241…41242
Fsetpoint
sets the required temperature of the heater in °C. Note that the heater has no cooling functionality; a setpoint
lower than the actual heater temperature effectively switches the heater off.
Capacity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
R
200
21
1/13
0x8168…0x8169/33129…33130
This parameter returns the maximum readout and control value (100%), expressed in units corresponding to parameter
Capacity Unit
. For the CEM control module, this parameter is fixed to 200 (°C); it might come in handy when scaling the value
of parameter
Measure
(1 °C = 32000/200).
Capacity Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[7]
R
n/a
129
1/31
0x81F8…0x81FB/33273…33276
This parameter returns the capacity unit in which
Fmeasure
and
Fsetpoint
are expressed. For the CEM control module, this
parameter is fixed to °C.