Brewer
FLEX
Access
Exam Table
Printed in USA © 2015
4
Document # 2102861 RevA
Power Cord
• Length .....................................Extends 70 in. from Table
• Type ..............................................................Detachable
• Conductors .................................. 18 AWG / 3 Conductor
• Grade ...............................................Hospital Grade Plug
Type of protection against electrical shock
• Model 5700 and 5800 ............. Class 2 Double Insulated
• Models 5701 and 5801 ....................... Class 1 Grounded
• Patient applied parts (all models) .........................Type B
Type of protection against ingress of water
……Ordinary
Paper Roll Sizes
• Regular ......................................................18 in. x 3.5 in.
• Long ..........................................................21 in. x 3.5 in.
Certifications* ......
UL60601-1;CAN-CSA C22.2 No. 601.1
Operating Conditions
• Temperature Range ......................................65° to 85°F
• Relative Humidity .........................................10% to 90%
Transportation and Storage
• Temperature .............................................-20°F to 150°F
• Humidity ........................................................10% to 90%
* Classified in the United States and Canada per
the standards listed above.
Weight of Table
..................................................... 315 lbs
Load Rating (maximum)
• Seat/Back 5700/5701/5800/5801 ....................... 700 lbs.
• Leg Support ......................................................... 50 lbs.
• Pass Through Work Surface (optional) ................ 15 lbs.
• Grab Bars (optional) ........................................... 100 lbs.
Range of Motion
• Back Section Range ..................0° (horizontal) to 75-85°
• Table Top Height Range ..........18 ± 0.5 in. to 37 ± 1.0 in.
• Pelvic Tilt Range
(Models 5701 & 5801 Only) ......0° (horizontal) to 5° (up)
Upholstery Dimensions
• Upholstered Top ............... 24 to 28 in. wide x 60 in. long
• Upholstered Leg Support… ....…18 in. long x 18 in. wide
• Overall Length of Upholstery .......................... 60 in. long
• Upholstery with Leg Support Extended .......... 78 in. long
Electrical Rating
• 5700/5800 ............ . 115 VAC nominal, 60 HZ, 6.0 amps
• 5701/5801 W/Outlet 115 VAC nominal, 60 HZ, 6.0 amps
• Fuse Rating:
• Control Box Fuse Rating ................................ 250V/2.5A
• Option/Warmer & Outlet Fuse Rating ............. 250V/3.0A
• Outlet Max Load .......................................................2.0A
Duty Cycle (10%)
.................... 2 minutes on / 18 minutes off (motor run time)
SPECIFICATIONS
Certifications
This product has been evaluated with respect to electrical
shock, fire, and mechanical hazards only in accordance
with UL60601-1; CAN/CSA C22.2 No. 601.1
NOTICE
Electromagnetic Interference (EMI)
This product is designed and built to minimize
electromagnetic interference with other devices; however,
if interference is noticed between another device and this
product, remove the interfering device from the room or
plug this product into an isolated circuit.
NOTICE
Model Configurations
Model
Power Back
Pneumatic
Manual Back
Pelvic Tilt
Drawer
Warmer
Return To
Chair
Outlet
5700
X
5701
X
X
X
X
5800
X
X
5801
X
X
X
X
X