Figure 41. Pelvic Tilt Adjustment
MAINTENANCE
SERVICE
For the latest in Cleaning Guidelines visit:
www.brewercompany.com
Table Care
Loose Fasteners and Damaged Components Hazard
Failure to perform periodic inspections of the
table could result in personal injury or equipment
damage.
CAUTION
More Cleaning Instructions Available
For the latest in Cleaning Guidelines visit:
www.brewercompany.com
NOTICE
Preventative Maintenance
• Inspect table every 6 months.
• Periodically inspect the electrical and foot control
cords to ensure they are free of cuts or damage.
• Inspect the mechanical functions to ensure satisfac-
tory operation.
• Check fasteners to make sure they are present and
tightened securely.
• Lubricate the backrest hinge with a food grade
silicon lubricant (Figure 41). Cover surrounding
area with cloth to protect from overspray and wipe
excess.
WARNING
Shock Hazard
Disconnect power by removing
plug from the wall receptacle
before changing fuses.
Fuse Cap
Figure 42.
Drawer Warmer Fuse Replacement
(Models 5701 and 5801)
1. Disconnect power cord.
2. Remove fuse cap from back of table near the power
cord (Figure 42).
a. Insert flat bladed screwdriver into slot.
b. Push in and turn counterclockwise 1/4 turn.
c. Pull cap/fuse from holder.
3. Replace fuse. Fuse type is 5 x 20mm, 250V/3.0A.
a. Insert replacement fuse in cap.
b. Insert cap/fuse into holder.
c. Push and twist 1/4 turn clockwise with screwdriver
to lock.
4. Connect power cord and check drawer warmer for
proper operation.