67XXU USB/Serial Scale Service Manual
49
HOLD
AD.H.SP
YES
NO
Enable or disable use of the
high speed A/D converter after
entering the HOLD mode
NO
HLD.MD
NONE
auto
Hold mode
(1) NONE= off
PS.PEK
(2) PS.PEK=Positive Peak number
Hold mode. Scale will display and
refresh the positive peak value from
last zero setting.
NG.PEK
(3) NG.PEK=Negative PEAK
number Hold mode. It is similar to
PS.PEK, but a negative number is
used.
TOGLE
(4) TOGLE=Press [HOLD] key to
enter HOLD mode. Stores the
display value. Will stay on the
display for the duration of time set
at HLD.TM.
AVERG
(5) AVERG= Average HOLD mode.
Time it looks at the weight before
updating the display. Set by
AVG.TM. Weight or vibration must
fall within window of HLD.RG also.
AUTO
(6) AUTO=Auto hold mode. It is
similar to AVERAG mode, but will
allow more weight to be added.
AVG.TM
1-60
3
Average data time for HOLD
mode
1-60 seconds
STB.TM
3XAVG.TIM -
255
3X
AVG.TIM
Wait time for the scale to be
stable in HOLD mode
Window which has to be 3 times
larger than AVG.TM
HLD.TM
0-65535
0
Data HOLD time
0=data will be frozen until HOLD
key pressed.
1-65535=data frozen time is 1-
65535s, after the time elapses,
scale will exit HOLD mode.
HLD.RG
0-255
5
Hold range
Window of vibration for Auto to
work.
0=any data can be averaged
1-255= only the data which
vibration is in 1-255d can be
averaged and held.
hd.key
hold
hold
HOLD/PRINT key
(1) HOLD= only HOLD key
(2) PRINT= only PRINT key
hld.pr
(3) HLD.PR= HOLD and PRINT key
USER
SubMenu1 SubMenu2
Option
Default
Parameter Description
Comment