6.2 Geographical Adjustment
1.
When GEO is selected, press the
[TARE]
key to enter Geographical
Adjustment.
2.
When
CODE
is displayed, press the
[UNIT
] key to select geographical position
code or input user local gravity value directly.
3.
When CODE is selected, choose the position code of the scale being used
(00-70) according to the elevation and latitude from
by using the
[HOLD]
and
[UNIT]
keys.
4.
Press
[TARE]
key to confirm.
CAUTION! Only an authorized manufacturer representative or certified
verification personnel may make these changes. Changing the Geographical
setting alters the calibration values!