32
67XXU USB/Serial Scale Service Manual
When scale first enters this working mode, it will display the largest positive number
that is from the time of zero-point set. The scale also will always catch the largest
positive number and display it.
To exit HOLD mode, press the
[HOLD]
key again.
Negative Peak HOLD
When USER-HOLD-HLD.MD is set to
NG.PEK
, the hold mode is in negative peak hold
mode.
When the scale first enters this mode, it will display the largest negative number from
the time of zero-point set. The scale also will always catch the largest negative number
and display it.
To exit HOLD mode, press the
[HOLD]
key again.
Toggle HOLD
When USER-HOLD-HLD.MD is set to
TOGLE
, the hold mode is toggled between
HOLD and MANUAL HOLD function.
After entering this mode, the scale will freeze and display a number if scale is stable.
Only the weight that is over USER-OTHER-NLD.RG (zero 'dead' band) can be held.
To exit HOLD mode, press the
[HOLD]
key again. If the time of the scale being unstable
is more than USER-HOLD-STB.TM,
STB.ER
will be shown.
Press the
[TARE]
key to start averaging again or press the
[HOLD]
key to exit.
Average HOLD
Set the USER-HOLD-HLD.MD parameter to
AVERG
.
After entering this mode the scale will freeze and display number if the scale is stable.
If the scale is not stable, but the variation is less than the USER-HOLD-DYN.RG
setting, the scale will average data in USER-HOLD-AVG.TM, then freeze and display
the number. Only the weight that is over USER-OTHER-NLD.RG can be frozen.
The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the
time of scale variation is over USER-OTHER-NLD.RG and is more than USER-HOLD-
STB.TM,
STB.ER
will be shown.
Press
[UNIT]
or
[TARE]
to start averaging again, or press the
[HOLD]
key to exit.
Auto HOLD
Set the USER-HOLD-HLD.MD parameter to
AUTO
. Different items can be weighed
one after another without pressing any buttons.
After entering this working mode, scale will freeze and display number if scale is stable.
Only the weight that is over USER-OTHER-NLD.RG can be frozen. If the held weight
is moved away and a new load is put on the scale, the new load will automatically be
held.
The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the
time of scale variation being over USER-OTHER-NLD.RG is more than USER-HOLD-
STB.TM,
STB.ER
will be shown. Press the
[TARE]
key to start averaging again or press
the
[HOLD]
key to exit.