Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment—
guidance
Radiated
RF (iEC
61000-4-3)
3 V/m
80 Mhz to
2.5 GHz
10 V/m
80 Mhz to
1000 Mhz
h
3 V/m
1 ghz to
2.7 GHz
i
8 V/m
5.1 GHz to
5.8 GHz
d = 0.5 √P (80 MHz to 1000
MHz)
d = 0.29 √P (380 MHz to 390
MHz)
d = 0.36 √P (430 MHz to 470
MHz)
d = 0.21 √P (800 MHz to 960
MHz)
d = 1.7 √P (1 GHz to 2.7 GHz)
d = 0.21 √P (1.7 GHz to 1.99
GHz)
d = 0.21 √P (2.4 GHz to 2.57
GHz)
d = 0.63 √P (5.1 GHz to 5.8 GHz)
where
P
is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and
d
is the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,
f
should be less
than the compliance level
in each frequency range.
g
interference may occur in
the vicinity of equipment
marked with the following
symbol:
non-iodizing
electromagnetic
radiation
a.
The lATiTUDE Communicator is intended for use in the electromagnetic
environment specified above. The customer or the user of the LATITUDE
Communicator should assure that it is used in such an environment.
b. At 80 MHz and 1 GHz, the higher frequency range applies.
Содержание LATITUDE 6290
Страница 1: ...LATITUDE Communicator Comunicador LATITUDE PATIENT MANUAL MANUAL DEL PACIENTE...
Страница 2: ......
Страница 3: ...English 1 Espa ol 91...
Страница 4: ......
Страница 5: ...1 LATITUDE Communicator Patient Manual English LATITUDE Communicator English...
Страница 10: ...6 LATITUDE Communicator Patient Manual English...
Страница 94: ...90 LATITUDE Communicator Patient Manual English...