80
LATITUDE
®
Communicator Patient Manual
English
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment—guidance
Voltage
dips, short
interruptions,
and voltage
variations on
power supply
input lines
(iEC
61000-4-11)
<5%
U
T
b
(>95% dip
in
U
T
) for 0.5
cycle
<5%
U
T
(>95% dip
in
U
T
) for 0.5
cycle
Mains power quality
should be that of a
typical commercial
or hospital
environment.
40%
U
T
(60% dip
in
U
T
) for 5
cycles
40%
U
T
(60% dip
in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for 25
cycles
70%
U
T
(30% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for 5 sec
<5%
U
T
(>95% dip in
U
T
) for 5 sec
Power
frequency
(50/60 Hz)
magnetic field
(iEC
61000-4-8)
3 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
a.
The lATiTUDE Communicator is intended for use in the
electromagnetic environment specified above. The customer or
the user of the lATiTUDE Communicator should assure that it is
used in such an environment.
b.
U
T
is the AC main’s voltage prior to application of the test level.
Содержание LATITUDE 6290
Страница 1: ...LATITUDE Communicator Comunicador LATITUDE PATIENT MANUAL MANUAL DEL PACIENTE...
Страница 2: ......
Страница 3: ...English 1 Espa ol 91...
Страница 4: ......
Страница 5: ...1 LATITUDE Communicator Patient Manual English LATITUDE Communicator English...
Страница 10: ...6 LATITUDE Communicator Patient Manual English...
Страница 94: ...90 LATITUDE Communicator Patient Manual English...