One Yellow Collecting Wave
Description:
The Communicator was unable to
start an interrogation of your implanted
device, or your implanted device
was out of range at the time of the
attempted interrogation.
Action:
•
Ensure the Communicator is
optimally placed as described
in “Where to Place Your
Communicator” on page 12.
• Face the Communicator. Sit directly
in front of the Communicator. Make
sure you are within 10 ft (3 m) of the
Communicator.
• Move any wireless electronic
products (such as cordless or
cellular phones or baby monitors)
that are within 3 ft (1 m) of the
Communicator.
To verify that troubleshooting was a
success:
•
Press the heart button to start
another interrogation. If the
interrogation was a success, all
three Collecting Waves will light
green for 2 minutes.