81
LATITUDE
®
Communicator Patient Manual
English
Table 3. Guidance and manufacturer’s declaration—
electromagnetic immunity—for equipment and systems that are
not life supporting
a b c d
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment—
guidance
Portable and mobile
RF communications
equipment should be used
no closer to any part of the
lATiTUDE Communicator,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distances:
Conducted
RF (iEC
61000-4-6)
3 Vrms
150 khz
to 80 Mhz
3 Vrms
10 Vrms in
isM and
amateur
radio bands
between
150 khz and
80 Mhz
e
d
= 1.7 √
P
(150 kHz to 80 MHz)
where
P
is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and
d
is the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,
f
should be less
than the compliance level
in each frequency range.
g
Содержание LATITUDE 6290
Страница 1: ...LATITUDE Communicator Comunicador LATITUDE PATIENT MANUAL MANUAL DEL PACIENTE...
Страница 2: ......
Страница 3: ...English 1 Espa ol 91...
Страница 4: ......
Страница 5: ...1 LATITUDE Communicator Patient Manual English LATITUDE Communicator English...
Страница 10: ...6 LATITUDE Communicator Patient Manual English...
Страница 94: ...90 LATITUDE Communicator Patient Manual English...