63
Troubleshoo
ting
LATITUDE™ Communicator
Patient Manual
◊ Two Yellow Collecting Waves
Issue:
The Communicator started but was not able
to complete the interrogation within the time
allowed.
Action:
•
Ensure the Communicator is optimally
placed as described in
.
•
Face the Communicator. Sit directly in front
of the Communicator. Make sure you are
within 1 m (3 ft) of the Communicator.
•
Remain still until the interrogation is
complete, as shown in the Verify section
on
page 64. Do not move away from the
Communicator.
•
Move any wireless electronic products
(such as cordless or cellular phones, baby
monitors, or modems/routers) at least 1 m
(3 ft) away from the Communicator.
• Press the Heart button to start another
interrogation. Refer to “Using the Heart
Button” on page 16 for the light
sequence when the Heart button is
pressed.